Assembling<br />

Mounting the expiration valve<br />

● The parts must be entirely dry to prevent malfunctioning.<br />

● Hold stopper by the flange and place diaphragm on the collar<br />

of the stopper. Be careful to fit the diaphragm properly.<br />

● Insert stopper with diaphragm on top into the housing from<br />

below and screw in tightly.<br />

If the expiration valve has an optional water trap:<br />

● Fit the collecting jar.<br />

Medicament nebuliser<br />

● Assemble in accordance with separate Instructions for Use,<br />

see page 92.<br />

Breathing gas humidifier<br />

● Assemble in accordance with separate Instructions for Use.<br />

Installation, see page 26 and page 28.<br />

Before Reusing on Patient<br />

● Assemble machine as described under "Preparing for use"<br />

on page 22.<br />

● Carry out checks to ensure readiness for operation, see<br />

"Device Check" on page 34.<br />

Preparing<br />

Assembling<br />

Before Reusing on Patient<br />

Instructions for Use <strong>EvitaXL</strong> 163<br />

058<br />

059<br />

060