Gas supply<br />

● Screw the connecting hoses for medical air (air) and<br />

oxygen (O2) from the piped supply into the back panel of<br />

<strong>EvitaXL</strong> and plug their connectors into the terminal units.<br />

The compressed gases must be dry and free from dust<br />

and oil. Gas pressure must be 3 to 6 bar.<br />

Connecting Evita Remote (optional)<br />

● The plug-in card for Evita Remote may only be installed and<br />

programmed by specialists.<br />

For remote control of <strong>EvitaXL</strong> via the Remote Pad for parallel,<br />

remote operation of the following LED and key functions:<br />

1 red LED – to indicate warning messages<br />

2 yellow LED – to indicate caution and advisory messages<br />

3 »g Alarm Silence« key – to suppress the alarm tone for<br />

approx. 2 minutes<br />

4 »Alarm Reset« key – to acknowledge alarm messages<br />

5 » Neb.« key – to start and end medicament<br />

nebulisation<br />

6 »O2 Y suction« key – for oxygenation for bronchial<br />

suctioning<br />

7 »Insp. hold« key – for sustained, manually induced inspiration<br />

8 »Exp. hold« key – for extended and sustained expiration<br />

The function of the respective LEDs and keys is the same as<br />

on <strong>EvitaXL</strong> and is described in the application chapters of<br />

these Instructions for Use.<br />

Preparing for use<br />

Gas supply<br />

Connecting Evita Remote (optional)<br />

Air O2<br />

Instructions for Use <strong>EvitaXL</strong> 31<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

085<br />

074