Using HME, bacterial filters and hose systems*<br />

Additional components in the breathing system or components<br />

which diverge from the standard hose system can considerably<br />

increase the inspiratory and expiratory breathing<br />

resistance and exceed standard requirements.<br />

Examples: Insp./exp. filters, HME**, coaxial hoses<br />

In general, the <strong>EvitaXL</strong> is designed to minimise the breathing<br />

effort made by the patient and is therefore not intended for the<br />

use of insp./exp bacterial filters.<br />

The use of bacterial filters or HMEs therefore requires particular<br />

care and monitoring by the user. Especially during medicament<br />

nebulisation and humidifying, the resistance of the<br />

expiratory filter may increase gradually.<br />

A higher breathing resistance leads to a greater breathing and<br />

trigger effort during assisted ventilation. Under unfavourable<br />

conditions, this can lead to an undesirable intrinsic PEEP. This<br />

can be recognised by the fact that the expiratory flow does not<br />

return to "0" at the end of expiration. If the PEEP is unacceptably<br />

high, it is indicated by the alarm »PEEP high !!!«. The current<br />

PEEP is then approx. 8 mbar above the set PEEP. Check<br />

and replace the bacterial filter or HME if they are the cause of<br />

the PEEP alarm.<br />

A breathing resistance in the patient connection cannot be<br />

monitored directly by the ventilator. For this reason:<br />

● Determine inspiratory and expiratory breathing resistance<br />

in the patient system before ventilation in standby by means<br />

of the device check.<br />

● Check the condition of the patient and the ventilator's measured<br />

values for volume and resistance frequently.<br />

● Observe the Instructions for Use for the HME, filter and<br />

coaxial hose systems in use.<br />

● Do not use an HME together with a medicament nebuliser<br />

or breathing gas humidifier. This can lead to a greater<br />

breathing resistance.<br />

* Only applies to hose systems which are not described in these<br />

Instructions for Use.<br />

** Heat moisture exchanger<br />

Preparing for use<br />

Using HME, bacterial filters and hose systems<br />

Instructions for Use <strong>EvitaXL</strong> 25