Technical Data<br />

Equipment outputs<br />

Analogue interface<br />

Voltage level 0 to 4.095 V<br />

Impedance of analogue channels The output impedance is 200 Ω.<br />

The input impedance of on-line equipment should be not less than 1 MΩ<br />

otherwise the output signal will be distorted.<br />

Accuracy 0 V : 0 V to 0.005 V<br />

4.095 V : 4.075 V to 4.115 V<br />

Electrical isolation The connecting socket is electrically isolated from the equipment electronics.<br />

The test voltage for electrical isolation is 1500 V. The analogue<br />

channels are not electrically isolated from one another.<br />

Pin assignment<br />

Pin 3<br />

Pin 1<br />

3<br />

Pin 2<br />

Medical appliance connector<br />

Instructions for Use <strong>EvitaXL</strong> 199<br />

<strong>EvitaXL</strong><br />

1<br />

2<br />

Connector housing<br />

GND<br />

Channel 1<br />

Channel 2