Evita 4 Link (optional)<br />
Evita 4 Link (optional)<br />
In addition to the standard RS 232 interface COM 1, <strong>EvitaXL</strong><br />
also has two serial RS 232 interfaces, COM 2 and COM 3,<br />
two CAN interfaces and an analogue interface with two<br />
channels.<br />
On both serial interfaces<br />
COM 2 and COM 3, the following can be output (optional):<br />
— The LUST protocol*,<br />
— the MEDIBUS protocol,<br />
— a printer protocol.<br />
The LUST protocol and printer protocol can only be output on<br />
one serial interface, while the MEDIBUS protocol can be output<br />
simultaneously on both serial interfaces.<br />
Other equipment, e.g. printers, may only be connected to<br />
the COM interfaces if <strong>EvitaXL</strong> is connected to the mains<br />
power supply via a mains power cable or if it has been<br />
earthed via the earth connection on the back of <strong>EvitaXL</strong>.<br />
Electric power may pose a hazard in all other cases.<br />
For output of measured values, status messages and alarm<br />
messages to on-line equipment for monitoring, documentation<br />
or processing.<br />
The on-line equipment may be a Dräger device or a device<br />
made by any other manufacturer.<br />
All data transmitted in this way are intended for information<br />
only and must not be used as the sole basis for deciding on a<br />
particular therapy.<br />
The RS 232 interfaces meet the requirements of the standards<br />
"EIA Standard RS 232 C" and "CCITT V.24".<br />
* For a detailed description of the LUST protocol, see "Technical<br />
data, LUST protocol" on page 201.<br />
178 Instructions for Use <strong>EvitaXL</strong>