For Your Safety and that of Your Patients<br />

Safe use of the equipment<br />

Do not use mobile telephones within 10 metres of<br />

ventilators!<br />

Mobile telephones may impair the functionality of electromedical<br />

equipment and endanger the patient * .<br />

General information on electromagnetic compatibility<br />

(EMC) according to the international EMC standard<br />

IEC 60601-1-2: 2001<br />

Medical electrical equipment needs special precautions<br />

regarding electromagnetic compatibility (EMC) and needs to<br />

be installed and put into service according to the EMC information<br />

provided in the technical documentation available<br />

from DrägerService upon request.<br />

Portable and mobile RF communications equipment can<br />

affect medical electrical equipment.<br />

Pins of connectors identified with the ESD warning<br />

symbol shall not be touched and not be connected<br />

unless ESD precautionary procedures are<br />

used. Such precautionary procedures may<br />

include antistatic clothing and shoes, the touch of a ground<br />

stud before and during connecting the pins or the use of<br />

electrically isolating and antistatic gloves. All staff involved in<br />

the above shall receive instruction in these procedures.<br />

* Dräger medical equipment meets the requirements for immunity to<br />

interference in accordance with the specific product standards and<br />

EN 60601-1-2 (IEC 60601-1-2).<br />

Depending on the type of mobile telephone used and on the application<br />

situation, however, field strengths exceeding the values<br />

specified in the applicable standards may develop in the immediate<br />

vicinity of the mobile telephone and therefore lead to faults and malfunctions.<br />

Adequate ventilation monitoring<br />

The following parameters are monitored by the built-in monitoring<br />

facilities of <strong>EvitaXL</strong>:<br />

— Airway pressure PAW<br />

— Expiratory minute volume MV<br />

— Inspiratory tidal volume, VTi<br />

— Inspiratory O2 concentration FiO2<br />

— Inspiratory breathing gas temperature T<br />

— Expiratory CO2 concentration, etCO2 (optional)<br />

— Apnoea time<br />

— Tachypnoea<br />

Changes in these parameters may be caused by:<br />

— Acute changes in the patient's condition<br />

— Incorrect settings and faulty handling<br />

— Equipment malfunctions<br />

— Failure of power and gas supplies<br />

If a fault occurs in this equipment, separate measuring instruments<br />

should be used.<br />

During the O2 Therapy, the monitoring functions of <strong>EvitaXL</strong> are<br />

restricted. Monitoring of SpO2 and the pulse is only available<br />

with the appropriate option.<br />

Back-up ventilation with an independent manual<br />

ventilation device<br />

If a fault is detected in <strong>EvitaXL</strong> so that its life-support functions<br />

are no longer assured, ventilation using an independent ventilation<br />

device must be started without delay – if necessary with<br />

PEEP and/or an increased inspiratory O2 concentration<br />

(e.g. with the Resutator MR 100).<br />

6 Instructions for Use <strong>EvitaXL</strong>