Measurements<br />

Insp. O2 concentration during medicament nebulisation<br />

Insp. O2 concentration during medicament<br />

nebulisation<br />

Use only medicament nebuliser 84 12 935 (white central<br />

section).<br />

If other medicament nebulisers are used, considerable<br />

deviations may occur in the tidal volume and the inspiratory<br />

O2 concentration!<br />

To minimise the deviation from the set O2 concentration,<br />

<strong>EvitaXL</strong> uses a mixed gas to drive the medicament nebuliser.<br />

In adult ventilation, this mixed gas is generated by switching<br />

over between compressed gases (medical air and oxygen) in<br />

synchronisation with inspiration.<br />

In paediatric ventilation, the nebuliser is operated continuously,<br />

with medical air or oxygen in alternation. The drive gas of<br />

the medicament nebuliser therefore roughly corresponds to<br />

the set FiO2.<br />

The graph shows the possible deviations of the applied O2<br />

concentration as a function of the set FiO2 with a minimal<br />

inspiratory flow (15 L/min) in adult ventilation or at ventilation<br />

frequencies above 12 bpm in paediatric ventilation.<br />

100<br />

90<br />

80<br />

70<br />

60<br />

50<br />

40<br />

30<br />

20<br />

20<br />

set O2 concentration FiO2 Vol.%<br />

228 Instructions for Use <strong>EvitaXL</strong><br />

applied O2 concentration FiO2 Vol.%<br />

30 40 50 60 70 80 90 100<br />