O2 sensor calibration<br />

— After replacing the O2 sensor (wait<br />

15 minutes for the O2 sensor to<br />

warm up).<br />

— When the measured and set values<br />

diverge by more than 2 % by volume.<br />

The O2 sensor can be calibrated during<br />

ventilation.<br />

● Press the key » Sensor<br />

Parameter«.<br />

● Touch the screen key »O2«. <strong>EvitaXL</strong><br />

displays the menu »O2«.<br />

Start O2 calibration:<br />

● Touch the »Start« screen key. The<br />

key turns green and <strong>EvitaXL</strong> calibrates<br />

the O2 sensor.<br />

Screen display:<br />

O2 calibration active<br />

The screen key »Start« turns pale green<br />

when calibration is complete.<br />

Sensors<br />

O2 sensor calibration<br />

Instructions for Use <strong>EvitaXL</strong> 109<br />