<strong>EvitaXL</strong> ventilates the patient in the set ventilation mode with<br />

increased O2 concentrations: In adult mode, the O2 concentration<br />

is 100 % by volume. In paediatric mode, the enriched<br />

concentration is 25 % higher than the set O2 concentration.<br />

If PEEP is not set to more than 4 mbar, PEEP will be applied<br />

automatically at 4 mbar. This PEEP allows <strong>EvitaXL</strong> to detect<br />

any subsequent disconnection.<br />

The other ventilation parameters remain unaffected.<br />

Screen display:<br />

Initial oxygen enrichment 180 s<br />

The remaining time is counted down continuously.<br />

This initial oxygen enrichment lasts for a maximum of 180 seconds.<br />

During this time <strong>EvitaXL</strong> waits for a disconnection for suction.<br />

The oxygen enrichment program is terminated by <strong>EvitaXL</strong> if<br />

there is no disconnection after the 180 seconds have elapsed.<br />

After disconnection for suction<br />

<strong>EvitaXL</strong> delivers a minimal flow for the duration of suction in<br />

order to detect automatically the end of the disconnection<br />

phase. The time available for suction is displayed on the<br />

screen continuously in seconds (example):<br />

Execute suction and reconnect 120 s<br />

If suction is ended and the system is reconnected within the<br />

displayed time, <strong>EvitaXL</strong> terminates the disconnection phase.<br />

Automatic interruption of oxygen enrichment<br />

If there is still no reconnection after 120 seconds, the oxygen<br />

enrichment program is interrupted. All alarms are immediately<br />

reactivated. <strong>EvitaXL</strong> continues ventilating in the set ventilation<br />

mode.<br />

After reconnection<br />

After reconnection, <strong>EvitaXL</strong> continues ventilating in the set ventilation<br />

mode, except that for 120 seconds an O2 concentration<br />

of 100 % by volume for adults or 25 % above the set<br />

O2 concentration for paediatric ventilation will continue to<br />

be delivered for final oxygen enrichment.<br />

Screen display:<br />

Final oxygen enrichment 120 s<br />

The remaining time is counted down continuously.<br />

If oxygen enrichment is to be interrupted:<br />

● Touch the screen key »O2 Y suction«.<br />

To view additional text information on oxygen enrichment:<br />

● Touch the screen key »? q «.<br />

To quit the menu:<br />

● Touch the screen key »x«.<br />

Additional functions<br />

Oxygen enrichment for bronchial suction<br />

Instructions for Use <strong>EvitaXL</strong> 95