Selecting Standby Mode<br />

— To perform the device check.<br />

— To keep <strong>EvitaXL</strong> ready for operation<br />

while the patient is absent.<br />

— To change the patient mode.<br />

— For the O2 therapy (optional).<br />

Ventilation does not take place in<br />

standby mode!<br />

● Press the key »O Start/Standby«<br />

and keep it depressed for 3 seconds.<br />

The ventilator is now in standby mode.<br />

Or<br />

● Press the key »O Start/Standby«.<br />

<strong>EvitaXL</strong> displays the menu »Start/<br />

Standby«.<br />

● Touch screen key »Standby«.<br />

● Confirm = press the rotary knob, the<br />

key turns green.<br />

1 Touch the screen key »Alarm Reset«<br />

in the field for alarm messages at the<br />

top of the screen.<br />

● Confirm = press the rotary knob, the<br />

key turns green.<br />

The ventilator is now in standby mode.<br />

If the patient mode or ideal body weight<br />

is changed in standby mode, new values<br />

will be calculated by <strong>EvitaXL</strong> for starting<br />

ventilation, see page 47.<br />

Selecting Standby Mode<br />

Instructions for Use <strong>EvitaXL</strong> 119<br />

➀<br />

202