22.11.2012 Views

EvitaXL

EvitaXL

EvitaXL

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Select printer protocol<br />

● See "Set interface" on page 136.<br />

In addition to automatically triggered printouts at predetermined<br />

intervals, printing can also be started manually via the<br />

on-screen function key »Print«, see "Configuration", "Define<br />

function keys", page 130.<br />

Analogue interface<br />

The analogue interface on <strong>EvitaXL</strong> has two channels to which<br />

measured value signals can be assigned as desired.<br />

Characteristics and PIN assignment<br />

See "Technical data", page 198.<br />

External voltages must not be applied here!<br />

Assign channels:<br />

● See "Set interface" on page 136.<br />

Measured<br />

value<br />

signal<br />

Designation Range/voltage<br />

level<br />

PAW Airway pressure –10 to 100 mbar<br />

0 to 4.095 V<br />

Flow Expiratory and inspiratory<br />

flow<br />

V Expiratory and inspiratory<br />

volume<br />

–196 to 196 L/min<br />

0 to 4.095 V<br />

0 to 2 L<br />

0 to 4.095 V<br />

MV Minute volume 0 to 41 L/min<br />

0 to 4.095 V<br />

f Ventilation frequency 0 to 150 bpm<br />

0 to 4.095 V<br />

FiO2 Inspiratory O2 concentration 0 to 100 Vol.%<br />

0 to 4.095 V<br />

R Resistance 0 to 100 mbar/L/s<br />

0 to 4.095 V<br />

C Compliance 0 to 250 mL/mbar<br />

0 to 4.095 V<br />

CO2 Expiratory O2 concentration 0 to 15 kPa<br />

0 to 4.095 V<br />

etCO2 End-expiratory CO2<br />

concentration<br />

0 to 15 kPa<br />

0 to 4.095 V<br />

NO Inspiratory flow for NOdomo 0 to 125 L/min<br />

0 to 4.095 V<br />

Evita 4 Link (optional)<br />

Preparation<br />

Instructions for Use <strong>EvitaXL</strong> 181