Calibrating the CO2 sensor<br />

— If the check values are not met on<br />

checking the calibration using calibration<br />

gas.<br />

Calibration gas containing N2O must<br />

not be used!<br />

● Switch on <strong>EvitaXL</strong> and wait at least<br />

3 minutes for the CO2 sensor to complete<br />

its warm-up phase.<br />

● First perform CO2 zero calibration,<br />

page 111, then:<br />

In the menu »CO2«:<br />

● Touch the screen key »Calibration«.<br />

<strong>EvitaXL</strong> displays the »Calibration« menu.<br />

● Connect the calibration gas supply.<br />

Use the cuvette from the calibration<br />

set!<br />

1 Connect the calibration gas cylinder<br />

and the cuvette of the calibration set<br />

to the hose.<br />

2 Remove the CO2 sensor from its<br />

park bracket and fit it to the cuvette of<br />

the calibration set.<br />

3 Read the CO2 and O2 concentration<br />

(if applicable) of the calibration gas<br />

from the calibration gas cylinder.<br />

Enter these concentration values with<br />

the screen knobs.<br />

Sensors<br />

Zero/check/calibrate CO2 sensor<br />

Instructions for Use <strong>EvitaXL</strong> 115<br />

2<br />

1<br />

3<br />

140<br />

048