Setting Ventilation Modes<br />
SIMV, SIMV/ASB<br />
SIMV, SIMV/ASB<br />
Synchronized Intermittent Mandatory<br />
Ventilation*<br />
Assisted Spontaneous Breathing**<br />
Combines mechanical (volume-controlled)<br />
ventilation with spontaneous breathing.<br />
The patient can breathe<br />
spontaneously between the mandatory<br />
ventilation strokes, contributing to the<br />
total minute volume. The spontaneous<br />
breaths can be supported with ASB.<br />
Mandatory ventilation strokes ensure a<br />
minimum level of ventilation in the times<br />
in between. This minimum ventilation is<br />
determined via the two set parameters<br />
tidal volume »VT« and frequency »f« and<br />
is the product of VT x f.<br />
The frequency can be reduced to zero<br />
during the weaning process. The device<br />
automatically changes to the ventilation<br />
mode CPAP or CPAP/ASB. This ventilation<br />
mode is also indicated on the<br />
screen. The screen key »SIMV« and the<br />
screen knobs for setting the SIMV<br />
parameters remain on display.<br />
Set the pattern of ventilation for SIMV<br />
and SIMV/ASB with the ventilation<br />
parameters:<br />
Tidal volume »VT«<br />
Insp. Flow »Flow«<br />
Frequency »f«<br />
Inspiration time »Tinsp«<br />
O2 concentration »O2«<br />
Positive end-expiratory pressure »PEEP«<br />
Pressure support »PASB«<br />
Pressure rise time »Ramp«<br />
--------------------------------------------------------------------------------<br />
To set:<br />
● Touch the appropriate screen knob.<br />
● Adjust to the desired value<br />
= turn rotary knob.<br />
● Confirm setting = press rotary knob.<br />
--------------------------------------------------------------------------------<br />
* For a detailed description of SIMV, please<br />
refer to page 211.<br />
** For a detailed description of ASB, please<br />
refer to page 216.<br />
PAW<br />
Flow<br />
Tinsp<br />
Pmax<br />
(configurable)<br />
50 Instructions for Use <strong>EvitaXL</strong><br />
PEEP<br />
1<br />
f<br />
Insp. flow<br />
fast<br />
rise time<br />
ASB pressure<br />
assist<br />
Trigger window<br />
slow<br />
rise time<br />
t<br />
t<br />
006<br />
108