Sensors<br />

Zero/check/calibrate CO2 sensor<br />

● Connect the calibration gas supply.<br />

Use the cuvette from the calibration<br />

set!<br />

1 Connect the calibration gas cylinder<br />

and the cuvette of the calibration set<br />

to the hose.<br />

2 Remove the CO2 sensor from its<br />

park bracket and fit it to the cuvette of<br />

the calibration set.<br />

3 Read the CO2 and O2 concentration<br />

(if applicable) of the calibration gas<br />

from the calibration gas cylinder.<br />

Enter these concentration values with<br />

the screen knobs.<br />

● Touch the screen knob.<br />

Enter the concentration<br />

= turn rotary knob,<br />

confirm = press rotary knob.<br />

If the calibration gas comprises CO2, O2<br />

and N2:<br />

● Enter the O2 concentration read off.<br />

If the calibration gas only comprises<br />

CO2 und N2:<br />

● Set the O2 concentration to »0«.<br />

● Touch the screen key »Gas Check«.<br />

<strong>EvitaXL</strong> displays the CO2 concentration<br />

FCO2 in the menu.<br />

Example: FCO2 5.0 Vol.%.<br />

After about 10 seconds, the value of<br />

FCO2 must match to within ±0.2 Vol.%<br />

the CO2 content of the calibration gas.<br />

If the calibration value is outside the permitted<br />

tolerance, the CO2 sensor must<br />

be recalibrated with test gas.<br />

● Push CO2 sensor back on the<br />

cuvette.<br />

114 Instructions for Use <strong>EvitaXL</strong><br />