EvitaXL
EvitaXL
EvitaXL
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Preparing<br />
Care list for the intensive-care ventilator <strong>EvitaXL</strong><br />
Care list for the intensive-care ventilator<br />
<strong>EvitaXL</strong><br />
The following list applies to non-infectious patients.<br />
If the ventilator is used with infectious patients, all parts in<br />
contact with breathing gas must be sterilised in addition to<br />
disinfection and cleaning.<br />
The parts in contact with breathing gas and listed below can<br />
be sterilised in hot steam at 134 o C. Refer to the column<br />
"Sterilise".<br />
The table is merely intended as a approximate guide. The<br />
instructions of the hospital's hygiene officer shall prevail!<br />
What How often How<br />
Components which<br />
can be conditioned<br />
<strong>EvitaXL</strong><br />
basic unit<br />
Trolley<br />
Hinged arm<br />
Compressed gas<br />
hose<br />
Ventilation hoses<br />
Y piece<br />
Water traps<br />
Collecting jars<br />
Recommended<br />
conditioning<br />
intervals<br />
Cleaning and disinfection<br />
machine,<br />
93 o C, 10 minutes<br />
Disinfect and clean Sterilise<br />
Wipe Immerse Hot steam<br />
134 o C,<br />
10 minutes<br />
Per patient no Outside no no<br />
Per patient no Outside no no<br />
Per patient/<br />
weekly<br />
yes no Possible yes<br />
Expiration valve Per patient/<br />
weekly *<br />
yes no Possible yes<br />
Spirolog flow sensor daily no **<br />
Outside Possible **<br />
no<br />
SpiroLife flow sensor daily no **<br />
Outside Possible **<br />
yes<br />
Temperature sensor daily no yes no yes<br />
CO2 sensor (optional) daily no yes ***<br />
no no<br />
Cuvette of the CO2<br />
sensor (optional)<br />
daily yes no yes yes<br />
Test filter for CO2<br />
sensor (optional)<br />
daily no yes *** no no<br />
Humidifier Per patient/weekly In accordance with separate Instructions for Use<br />
Medicament nebuliser<br />
(optional)<br />
Per patient/weekly In accordance with separate Instructions for Use<br />
Bacterial filter In accordance with separate Instructions for Use<br />
* Nebulisation may lead to formation of more extensive deposits necessitating more frequent replacement.<br />
** Special treatment, immerse in a disinfectant bath with 70 % ethanol, see page 158.<br />
*** Wipe-disinfect, e.g. with 70 % ethanol, see page 160.<br />
162 Instructions for Use <strong>EvitaXL</strong>