Optional operating concept<br />

QuickSet<br />

Optional operating<br />

concept<br />

QuickSet*<br />

Direct adjustment of PEEP or Pinsp<br />

Direct adjustment of PEEP is possible in<br />

all ventilation modes.<br />

Direct adjustment of Pinsp is possible in<br />

BIPAP and BIPAPAssist.<br />

● Touch the screen key »PEEP« or<br />

»Pinsp« in the respective ventilation<br />

mode.<br />

● Depress rotary knob and hold down<br />

for 3 secs.<br />

The adjuster changes to green with a yellow<br />

edge. Direct adjustment is active.<br />

● Keep rotary knob depressed and turn<br />

it. Set »PEEP« or »Pinsp«.<br />

The set value is now effective.<br />

On releasing the rotary knob, »PEEP« or<br />

»Pinsp« can still be set directly:<br />

● Depress again the rotary knob and<br />

turn it.<br />

Exceeding the standard setting range:<br />

● Release rotary knob for a short<br />

moment; the standard setting range<br />

can be exceeded.<br />

Set »PEEP« or »Pinsp«:<br />

● Depress again the rotary knob and<br />

turn it.<br />

* Lung Protection Package option<br />

18 Instructions for Use <strong>EvitaXL</strong><br />