Operation with mains power<br />
The ventilator switches to operation with mains power when<br />
connected to the mains power supply. The integrated battery is<br />
recharged during mains power operation. The external battery<br />
is recharged once the integrated battery is fully charged.<br />
The LED next to the plug symbol s lights up yellow.<br />
If the mains power supply fails, the ventilator automatically<br />
switches over to the external battery.<br />
If an external battery is not available in the event of a power failure,<br />
the ventilator switches over to the integrated battery and<br />
continues operation for at least 10 minutes (provided that the<br />
integrated battery was fully charged).<br />
● The mains power supply must be restored without delay.<br />
To ensure that the battery is always fully charged:<br />
● Connect the ventilator to the mains power supply and<br />
switch it on.<br />
The ventilator may only remain connected to the mains<br />
supply in well ventilated rooms.<br />
Oxyhydrogen gas is generated when charging batteries<br />
and may explode when a sufficient concentration has<br />
accumulated.<br />
● Leave the ventilator in standby mode or start ventilation.<br />
Operation with integrated battery<br />
Mains power/DC power operation<br />
Operation with mains power<br />
Operation with integrated battery<br />
The ventilator immediately switches over to operation with the<br />
integrated battery if the mains power supply fails without an<br />
external battery being connected or if the external battery is<br />
exhausted.<br />
The integrated battery is merely intended for emergency<br />
use and not for normal operation.<br />
The LED next to the symbol for the integrated battery lights up<br />
yellow.<br />
The green LED next to the symbol for the integrated battery<br />
goes out, as it is no longer fully charged.<br />
When <strong>EvitaXL</strong> switches over to the integrated battery, the following<br />
advisory message is displayed:<br />
Int. battery in operation !<br />
The time available for operation with the integrated battery<br />
depends on its charge state. If the battery is fully charged,<br />
operation can continue for at least 10 minutes.<br />
A caution message is displayed after 8 minutes of operation:<br />
Int. battery only 2 minutes left !!<br />
● The ventilator must be reconnected to the mains power supply<br />
within 2 minutes<br />
or<br />
● it must be connected to a fully charged external battery!<br />
After an operating time of 10 minutes has elapsed, the following<br />
alarm message is displayed:<br />
Int. battery discharged !!!<br />
● The power supply must be restored immediately, either<br />
from the mains supply or from a fully charged external battery,<br />
otherwise ventilation will be interrupted.<br />
After being powered by the integrated battery:<br />
● Recharge the integrated and external batteries as soon as<br />
possible, see "Charging the batteries", page 171.<br />
Instructions for Use <strong>EvitaXL</strong> 175