Operation with mains power<br />

The ventilator switches to operation with mains power when<br />

connected to the mains power supply. The integrated battery is<br />

recharged during mains power operation. The external battery<br />

is recharged once the integrated battery is fully charged.<br />

The LED next to the plug symbol s lights up yellow.<br />

If the mains power supply fails, the ventilator automatically<br />

switches over to the external battery.<br />

If an external battery is not available in the event of a power failure,<br />

the ventilator switches over to the integrated battery and<br />

continues operation for at least 10 minutes (provided that the<br />

integrated battery was fully charged).<br />

● The mains power supply must be restored without delay.<br />

To ensure that the battery is always fully charged:<br />

● Connect the ventilator to the mains power supply and<br />

switch it on.<br />

The ventilator may only remain connected to the mains<br />

supply in well ventilated rooms.<br />

Oxyhydrogen gas is generated when charging batteries<br />

and may explode when a sufficient concentration has<br />

accumulated.<br />

● Leave the ventilator in standby mode or start ventilation.<br />

Operation with integrated battery<br />

Mains power/DC power operation<br />

Operation with mains power<br />

Operation with integrated battery<br />

The ventilator immediately switches over to operation with the<br />

integrated battery if the mains power supply fails without an<br />

external battery being connected or if the external battery is<br />

exhausted.<br />

The integrated battery is merely intended for emergency<br />

use and not for normal operation.<br />

The LED next to the symbol for the integrated battery lights up<br />

yellow.<br />

The green LED next to the symbol for the integrated battery<br />

goes out, as it is no longer fully charged.<br />

When <strong>EvitaXL</strong> switches over to the integrated battery, the following<br />

advisory message is displayed:<br />

Int. battery in operation !<br />

The time available for operation with the integrated battery<br />

depends on its charge state. If the battery is fully charged,<br />

operation can continue for at least 10 minutes.<br />

A caution message is displayed after 8 minutes of operation:<br />

Int. battery only 2 minutes left !!<br />

● The ventilator must be reconnected to the mains power supply<br />

within 2 minutes<br />

or<br />

● it must be connected to a fully charged external battery!<br />

After an operating time of 10 minutes has elapsed, the following<br />

alarm message is displayed:<br />

Int. battery discharged !!!<br />

● The power supply must be restored immediately, either<br />

from the mains supply or from a fully charged external battery,<br />

otherwise ventilation will be interrupted.<br />

After being powered by the integrated battery:<br />

● Recharge the integrated and external batteries as soon as<br />

possible, see "Charging the batteries", page 171.<br />

Instructions for Use <strong>EvitaXL</strong> 175