Setting alarm limits<br />

● Press key » _ Alarm Limits«.<br />

<strong>EvitaXL</strong> opens the menu »Alarm Limits«.<br />

> = upper alarm limit<br />

< = lower alarm limit<br />

The values for the upper and lower alarm<br />

limit shown in the screen keys are initial<br />

values which are effective whenever the<br />

ventilator is switched on. However, they<br />

can also be configured specifically as<br />

required by the hospital, see "Configuration",<br />

page 126.<br />

The actual measured value is displayed<br />

between the upper and lower alarm limits.<br />

To set:<br />

● Touch the required screen key,<br />

it turns yellow,<br />

set = turn rotary knob,<br />

confirm = press rotary knob.<br />

The alarm limits for the optional measured<br />

value etCO2 can be viewed by<br />

touching "Limits 2".<br />

● Touch the screen key »Limits 2«.<br />

To deactivate (MV < for example):<br />

● »Reduce »MV < « until the following<br />

advisory message is displayed:<br />

»MV < off? Press and turn<br />

rotary knob«<br />

● Confirm advisory message<br />

= press rotary knob.<br />

● Continue turning rotary knob until<br />

dashes (––) appear on the display.<br />

● Confirm = press rotary knob.<br />

Setting alarm limits<br />

Instructions for Use <strong>EvitaXL</strong> 79<br />