Replace cooling air filter<br />
— Clean if soiled or after 4 weeks at the latest.<br />
Replace after 1 year at the latest.<br />
1 Remove cooling air filter from its slot on the back of <strong>EvitaXL</strong>.<br />
● Fit a new cooling air filter or clean the installed filter in warm<br />
soapy water and dry thoroughly.<br />
● Insert the cooling air filter in the mount without creasing.<br />
● The old cooling air filter can be disposed with domestic<br />
waste.<br />
DC power supply MB:<br />
● On the back of <strong>EvitaXL</strong>, undo the two knurled screws securing<br />
the grating over the power supply fan and remove the<br />
grating.<br />
● Remove cooling air filter from its slot.<br />
● Fit a new cooling air filter or clean the installed filter in warm<br />
soapy water and dry thoroughly.<br />
● Insert the cooling air filter in the mount without creasing.<br />
● Refit the grating over the cooling air filter and tighten the<br />
knurled screws by hand.<br />
● The old cooling air filter can be disposed with domestic<br />
waste.<br />
Removing and reinserting ambient air filter<br />
— Clean after 4 weeks.<br />
Replace after 1 year at the latest.<br />
2 If necessary, swivel port to the left.<br />
3 Use coin to loosen screw, and remove protective cover.<br />
4 Remove the ambient air filter from the protective cover.<br />
● Slide the cleaned or new ambient air filter under the tabs.<br />
● Replace protective cover, and tighten screw with a coin.<br />
● Dispose of used ambient air filter with domestic waste.<br />
Replace cooling air filter<br />
Removing and reinserting ambient air filter<br />
Instructions for Use <strong>EvitaXL</strong> 167<br />
2<br />
4<br />
3<br />
1<br />
089<br />
217<br />
061<br />
062