Technical Data<br />

DC power supply<br />

DC power supply<br />

Electrical ratings for the DC socket<br />

DC input voltage 12 or 24 V battery<br />

Input current<br />

12 V battery typically 13 A, max. 30 A<br />

24 V battery typically 6 A, max. 15 A<br />

Performance characteristics<br />

Time bridged following a mains power failure<br />

(with fully operational batteries * )<br />

— Fully charged integrated battery typically 14 minutes; at least 10 minutes<br />

— Two fully charged external 12 V lead-gel batteries<br />

(18 43 303) with a capacity of 17 Ah each typically 2 hours<br />

External batteries **<br />

Type Lead-gel batteries, sealed maintenance-free<br />

Minimum capacity<br />

12 V battery 30 Ah (note battery charging current)<br />

24 V battery 15 Ah (note battery charging current)<br />

Max. charging current through the DC power supply (the<br />

battery used must be rated for at least this charging current)<br />

12 V battery approx. 8 A<br />

24 V battery approx. 4.5 A<br />

Charging time *** (for 18 43 303),<br />

2 batteries connected in series<br />

24 V lead-gel battery typically 8 to 12 hours<br />

Integrated batteries<br />

Type Lead-gel batteries, sealed maintenance-free<br />

Charging time ***<br />

DC power supply FB typically 1.5 to 2.5 hours<br />

DC power supply MB typically 2.5 to 4 hours<br />

Protection class I<br />

* see "Charge indication and battery charge state", page 172.<br />

** see "Mains power/DC power operation", page 170.<br />

*** see "Charging times", page 172.<br />