Technical Data<br />
DC power supply<br />
DC power supply<br />
Electrical ratings for the DC socket<br />
DC input voltage 12 or 24 V battery<br />
Input current<br />
12 V battery typically 13 A, max. 30 A<br />
24 V battery typically 6 A, max. 15 A<br />
Performance characteristics<br />
Time bridged following a mains power failure<br />
(with fully operational batteries * )<br />
— Fully charged integrated battery typically 14 minutes; at least 10 minutes<br />
— Two fully charged external 12 V lead-gel batteries<br />
(18 43 303) with a capacity of 17 Ah each typically 2 hours<br />
External batteries **<br />
Type Lead-gel batteries, sealed maintenance-free<br />
Minimum capacity<br />
12 V battery 30 Ah (note battery charging current)<br />
24 V battery 15 Ah (note battery charging current)<br />
Max. charging current through the DC power supply (the<br />
battery used must be rated for at least this charging current)<br />
12 V battery approx. 8 A<br />
24 V battery approx. 4.5 A<br />
Charging time *** (for 18 43 303),<br />
2 batteries connected in series<br />
24 V lead-gel battery typically 8 to 12 hours<br />
Integrated batteries<br />
Type Lead-gel batteries, sealed maintenance-free<br />
Charging time ***<br />
DC power supply FB typically 1.5 to 2.5 hours<br />
DC power supply MB typically 2.5 to 4 hours<br />
Protection class I<br />
* see "Charge indication and battery charge state", page 172.<br />
** see "Mains power/DC power operation", page 170.<br />
*** see "Charging times", page 172.<br />
200 Instructions for Use <strong>EvitaXL</strong>