Additional functions<br />

Oxygen enrichment for bronchial suction<br />

Oxygen enrichment for<br />

bronchial suction<br />

To avoid any risk of hypoxia during bronchial<br />

suction, <strong>EvitaXL</strong> offers a procedure<br />

for oxygen enrichment during the<br />

removal of secretions.<br />

After the program is started, <strong>EvitaXL</strong> ventilates<br />

the patient in the selected ventilation<br />

mode for an initial oxygen<br />

enrichment phase of 180 seconds.<br />

— In adult mode, the ventilator supplies<br />

100 % by volume O2, and in paediatric<br />

mode it delivers the set<br />

O2 concentration plus 25 % (for<br />

example: 60 % by vol. set; administered:<br />

75 % by vol.)<br />

When the ventilator is disconnected for<br />

suction, <strong>EvitaXL</strong> interrupts ventilation.<br />

During the suction phase, the audible<br />

alarms are suppressed so that the suction<br />

routine is not disturbed.<br />

After suction and automatically recognised<br />

reconnection, <strong>EvitaXL</strong> delivers an<br />

increased O2 concentration for the final<br />

oxygen enrichment phase of 120 seconds:<br />

— In adult mode, the O2 concentration<br />

is 100 % by volume. In paediatric<br />

mode, the enriched concentration is<br />

25 % higher than the set O2 concentration.<br />

During suction and for 2 minutes afterwards,<br />

the lower alarm limit for the<br />

minute volume is switched off. Other<br />

alarms are switched off during suction<br />

and for 15 seconds afterwards.<br />

Oxygen enrichment is only possible<br />

with a fully functioning flow sensor<br />

and if flow monitoring is switched on!<br />

Before suction<br />

● Touch the function key »Special Procedure...«.<br />

<strong>EvitaXL</strong> displays the menu<br />

»Additional Function«.<br />

● Touch the screen key »O2 Y suction«,<br />

the key turns yellow.<br />

● Confirm = press rotary knob. The key<br />

turns green and the oxygen enrichment<br />

program is started.<br />

94 Instructions for Use <strong>EvitaXL</strong><br />