Leak test<br />

This test must be performed:<br />

— after the device check,<br />

— after changing the hose system,<br />

— after changing the humidifier.<br />

In the checklist:<br />

● Touch the screen key »Airtight<br />

Check«.<br />

The result of the last leak test is displayed<br />

together with the following values:<br />

— Leakage<br />

— Compliance<br />

— Insp. Resistance<br />

— Exp. Resistance<br />

To start the leak test:<br />

● Touch the screen key »Check«.<br />

The actual leakage flow is displayed continuously<br />

throughout the test. A leakage<br />

flow of 300 mL/min at a pressure of<br />

60 mbar is permissible.<br />

After the leak test, the ventilator determines<br />

the compliance and the inspiratory<br />

and expiratory resistance of the<br />

hose system.<br />

The established compliance of the hose<br />

system is used by the ventilator for automatic<br />

correction of the volume-controlled<br />

ventilation strokes, as well as of the<br />

measured values for flow monitoring,<br />

see page 218.<br />

When changing the patient mode or<br />

type of humidifier, the ventilator automatically<br />

resets the values for hose compliance<br />

and hose resistance to the default<br />

values.<br />

To return to the start screen:<br />

● Touch the screen key »Start/<br />

Standby« in the »Start/Standby«<br />

menu.<br />

Preparing for use<br />

Leak test<br />

Instructions for Use <strong>EvitaXL</strong> 39<br />