As for ventilation hoses, water traps and associated jars,<br />

Y-piece, expiration valve (or, in the event of severe soiling,<br />

their individual parts), individual parts of the medicament<br />

nebuliser and adapter parts<br />

● Disinfect with moist heat (93 oC/10 minutes) in cleaning<br />

and disinfecting machine. Only use cleaning agent.<br />

If a cleaning and disinfection machine is not available:<br />

● Immerse the parts in a disinfectant bath, e.g. with Sekusept<br />

made by Messrs. Henkel.<br />

Comply with the manufacturer's instructions.<br />

● Then rinse with clean water, preferably from a soft water<br />

supply.<br />

● Shake water out thoroughly, and leave the products to dry.<br />

Expiration valve and its individual parts after disinfection<br />

● Rinse thoroughly with clear water, preferably from a soft<br />

water supply. Shake water out thoroughly.<br />

● After rinsing thoroughly, dry expiration valve.<br />

● After drying, sterilise in hot steam at 134 o C, otherwise liquid<br />

may remain in the pressure measuring line and impair<br />

correct functioning.<br />

Ventilation hoses, water traps and associated water jars,<br />

Y-piece, temperature sensor<br />

● These parts can be sterilised in hot steam at 134 o C.<br />

Bacterial filter<br />

● Must be conditioned in accordance with the separate<br />

Instructions for Use.<br />

Humidifier<br />

● Must be conditioned in accordance with the separate<br />

Instructions for Use.<br />

Preparing<br />

Disinfecting and Cleaning<br />