Additional functions<br />

Medicament nebulisation<br />

● Prepare the medicament nebuliser in accordance with its<br />

Instructions for Use.<br />

For use during adult ventilation<br />

1 Connect the nebuliser to the inspiratory side (temperature<br />

sensor side) of the Y-piece.<br />

2 Connect the inspiration hose to the medicament nebuliser.<br />

● Place the medicament nebuliser in the vertical position.<br />

● Using clamps, route the nebuliser hose back to the ventilator<br />

along the expiratory hose.<br />

For use during paediatric ventilation<br />

3 Insert the catheter connector (ISO cone Ø15/Ø11) in the<br />

inlet of the medicament nebuliser.<br />

4 Insert the adapter (ISO cone Ø 22/Ø11) in the outlet.<br />

5 Fit the corrugated hose (0.13 m long) to the outlet adapter.<br />

6 Remove the corrugated hose of the hose set from the<br />

inspiratory adapter of the Y-piece and connect it to the inlet<br />

adapter of the medicament nebuliser.<br />

● Connect the free end of the corrugated hose at the outlet<br />

of the medicament nebuliser to the inspiratory adapter of<br />

the Y-piece.<br />

7 Connect the nebuliser hose to the port on the front panel of<br />

the <strong>EvitaXL</strong>.<br />

● Fill the medicament nebuliser in accordance with the specific<br />

Instructions for Use.<br />

The effect of aerosols on sensors, filters and heat and<br />

moisture exchangers (HME) must be taken into account!<br />

Do not place a microbial filter on the nebuliser outlet during<br />

nebulisation!<br />

The measuring function of the flow sensor may be impaired.<br />

The flow resistance of filters is liable to increase and may<br />

impair ventilation.<br />

During medicament nebulisation, do not use a heat and<br />

moisture exchanger (HME) at the Y-piece. Risk of<br />

increased breathing resistance!<br />

92 Instructions for Use <strong>EvitaXL</strong><br />

5<br />

D<br />

6<br />

4<br />

1<br />

2<br />

1<br />

1<br />

1<br />

1<br />

1<br />

7<br />

2<br />

2<br />

2<br />

2<br />

2<br />

2<br />

3<br />

<strong>EvitaXL</strong><br />

037<br />

038<br />

039

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!