Message Cause Remedy<br />
Apnoea !!! Patient´s spontaneous breathing has<br />
stopped.<br />
Fault – Cause – Remedy<br />
Check condition of patient, if necessary<br />
apply controlled ventilation.<br />
Stenosis Check condition of patient.<br />
Check tube.<br />
Check hose system.<br />
Flow sensor not calibrated or faulty. Calibrate flow sensor.<br />
Replace if necessary.<br />
Apnoea alarm off ! Apnoea monitoring has been switched<br />
off in application mode NIV.<br />
Apnoea monitoring is not active when<br />
nebulising without flow monitoring.<br />
Apnoea ventilation !! Due to detected apnoea, the system has<br />
automatically switched over to mandatory<br />
ventilation.<br />
ASB > 4 s<br />
Not displayed in application<br />
mode »Mask/NIV«<br />
!!! The ASB phase was terminated<br />
3 consecutive times after 4 seconds.<br />
Set the upper alarm limit for apnoea<br />
monitoring to the required value again.<br />
Use external monitoring or switch on flow<br />
monitoring or stop nebulising.<br />
Check ventilation procedure. Return to the<br />
original ventilation mode by pressing the<br />
»Alarm Reset« key and confirm.<br />
Check condition of patient. Check tube.<br />
Test ventilation system for leaks.<br />
ASB > 1.5 s ! The ASB phase was terminated<br />
3 consecutive times after 1.5 seconds.<br />
Test ventilation system for leaks.<br />
ASB > Tinsp ! The ASB phase was terminated by a time<br />
limitation.<br />
Test ventilation system for leaks.<br />
Back-up ventilation !!! Tube blocked. Check tube.<br />
Breathing cycle not detected !!! The device does not deliver any gas. Set Pmax higher than PEEP setting.<br />
Extend alarm time TApnoea > or increase<br />
IPPV frequency.<br />
Device faulty. Disconnect patient from the device and<br />
continue ventilation without delay, using<br />
another independent ventilator.<br />
Call DrägerService.<br />
Check frequency ILV Slave<br />
Message on slave device<br />
! The frequency (breathing rate) of the<br />
master and slave device differs by more<br />
than 12 %.<br />
Check settings !! Power interruption while setting ventilation<br />
pattern or the alarm limits.<br />
Adjust the frequency of the slave device to<br />
that of the master device.<br />
Check pattern of ventilation and alarm<br />
limits. Acknowledge message by pressing<br />
»Alarm Reset« key and confirm.<br />
Clean CO2 cuvette !!! Cuvette window for the CO2 measurement<br />
is dirty.<br />
Use clean cuvette.<br />
Sensor window for the CO2 measurement<br />
is dirty.<br />
Clean CO2 sensor.<br />
CO2 measurement inop. !!! CO2 sensor faulty. Replace faulty CO2 sensor.<br />
CO2 measurement incorrect. The ventilator functions are not affected.<br />
Ensure adequate external monitoring<br />
without delay. Deactivate the internal<br />
CO2 monitoring.<br />
Call DrägerService.<br />
Instructions for Use <strong>EvitaXL</strong> 145