Message Cause Remedy<br />

Apnoea !!! Patient´s spontaneous breathing has<br />

stopped.<br />

Fault – Cause – Remedy<br />

Check condition of patient, if necessary<br />

apply controlled ventilation.<br />

Stenosis Check condition of patient.<br />

Check tube.<br />

Check hose system.<br />

Flow sensor not calibrated or faulty. Calibrate flow sensor.<br />

Replace if necessary.<br />

Apnoea alarm off ! Apnoea monitoring has been switched<br />

off in application mode NIV.<br />

Apnoea monitoring is not active when<br />

nebulising without flow monitoring.<br />

Apnoea ventilation !! Due to detected apnoea, the system has<br />

automatically switched over to mandatory<br />

ventilation.<br />

ASB > 4 s<br />

Not displayed in application<br />

mode »Mask/NIV«<br />

!!! The ASB phase was terminated<br />

3 consecutive times after 4 seconds.<br />

Set the upper alarm limit for apnoea<br />

monitoring to the required value again.<br />

Use external monitoring or switch on flow<br />

monitoring or stop nebulising.<br />

Check ventilation procedure. Return to the<br />

original ventilation mode by pressing the<br />

»Alarm Reset« key and confirm.<br />

Check condition of patient. Check tube.<br />

Test ventilation system for leaks.<br />

ASB > 1.5 s ! The ASB phase was terminated<br />

3 consecutive times after 1.5 seconds.<br />

Test ventilation system for leaks.<br />

ASB > Tinsp ! The ASB phase was terminated by a time<br />

limitation.<br />

Test ventilation system for leaks.<br />

Back-up ventilation !!! Tube blocked. Check tube.<br />

Breathing cycle not detected !!! The device does not deliver any gas. Set Pmax higher than PEEP setting.<br />

Extend alarm time TApnoea > or increase<br />

IPPV frequency.<br />

Device faulty. Disconnect patient from the device and<br />

continue ventilation without delay, using<br />

another independent ventilator.<br />

Call DrägerService.<br />

Check frequency ILV Slave<br />

Message on slave device<br />

! The frequency (breathing rate) of the<br />

master and slave device differs by more<br />

than 12 %.<br />

Check settings !! Power interruption while setting ventilation<br />

pattern or the alarm limits.<br />

Adjust the frequency of the slave device to<br />

that of the master device.<br />

Check pattern of ventilation and alarm<br />

limits. Acknowledge message by pressing<br />

»Alarm Reset« key and confirm.<br />

Clean CO2 cuvette !!! Cuvette window for the CO2 measurement<br />

is dirty.<br />

Use clean cuvette.<br />

Sensor window for the CO2 measurement<br />

is dirty.<br />

Clean CO2 sensor.<br />

CO2 measurement inop. !!! CO2 sensor faulty. Replace faulty CO2 sensor.<br />

CO2 measurement incorrect. The ventilator functions are not affected.<br />

Ensure adequate external monitoring<br />

without delay. Deactivate the internal<br />

CO2 monitoring.<br />

Call DrägerService.<br />

Instructions for Use <strong>EvitaXL</strong> 145