Switch on the medicament nebuliser:<br />

● Touch the function key »Special Procedure...«.<br />

<strong>EvitaXL</strong> displays the menu<br />

»Additional Function«.<br />

● Touch the screen key »<br />

Nebuliser«, the key turns yellow.<br />

● Confirm = press rotary knob. The key<br />

turns green and the nebuliser is operational.<br />

The message Nebuliser On !<br />

appears on the screen.<br />

Switch off the nebuliser:<br />

● Touch the screen key »<br />

Nebuliser«.<br />

The nebuliser is switched off automatically<br />

by the ventilator after 30 minutes.<br />

The flow sensor is automatically cleaned<br />

and calibrated after nebulisation.<br />

Screen display: Flow calibration<br />

● Remove remaining medicament.<br />

Follow the Instructions for Use of the<br />

medicament nebuliser.<br />

To view additional text information:<br />

● Touch the screen key »? q «.<br />

Additional functions<br />

Medicament nebulisation<br />

Instructions for Use <strong>EvitaXL</strong> 93<br />

134