Sensors<br />

Zero/check/calibrate CO2 sensor<br />

● Touch the screen knob.<br />

● Enter the<br />

concentration = turn rotary knob,<br />

confirm = press rotary knob.<br />

When using the standard calibration gas<br />

(5 Vol.% CO2 and 95 Vol.% N2):<br />

● Set O2 concentration to »0«,<br />

CO2 concentration to »5«.<br />

● Touch the »Start« screen key.<br />

During calibration, the following message<br />

is displayed on the screen:<br />

CO2 calibration. Please wait<br />

<strong>EvitaXL</strong> carries out calibration and<br />

confirms with the message:<br />

CO2 calibration ok<br />

Failed calibration is indicated by the ventilator<br />

with the message:<br />

CO2 calibration interrupted<br />

or<br />

CO2 calibration not ok<br />

● Repeat the calibration of the CO2<br />

sensor.<br />

If calibration still proves impossible, the<br />

CO2 concentration value entered may<br />

not be the same as that in the cylinder:<br />

● Check CO2 value entered,<br />

or<br />

calibration gas cylinder is empty:<br />

● Use a new calibration gas cylinder<br />

or<br />

sensor is defective:<br />

● Replace sensor.<br />

116 Instructions for Use <strong>EvitaXL</strong><br />