● Place the test filter in the<br />

CO2 sensor.<br />

<strong>EvitaXL</strong> displays the test value of the<br />

CO2 concentration FCO2 in the menu,<br />

example:<br />

FCO2 4.0 Vol.%<br />

This value must agree to within<br />

±0.3 Vol.% with the specification on the<br />

test filter.<br />

Example: 4.1 Vol.% on the filter:<br />

permitted value range: 3.8 to 4.4 Vol.%<br />

If the test value is outside the permitted<br />

tolerance, the test gas must be checked<br />

or calibrated.<br />

● Push CO2 sensor back on the<br />

cuvette.<br />

Checking CO2 calibration with calibration<br />

gas<br />

— if the specified calibration value<br />

was not met when testing with the<br />

test filter<br />

— at least once per half-year.<br />

Calibration gas containing N2O must<br />

not be used!<br />

● Switch on <strong>EvitaXL</strong> and wait at least<br />

3 minutes for the CO2 sensor to complete<br />

its warm-up phase.<br />

● First perform CO2 zero calibration,<br />

page 111, then:<br />

in the menu »CO2«:<br />

● Touch the screen key »Check<br />

sensor«.<br />

Sensors<br />

Zero/check/calibrate CO2 sensor<br />

Instructions for Use <strong>EvitaXL</strong> 113<br />