Preparing for use<br />

Power supply and connections<br />

Power supply and connections<br />

Electrical power supply<br />

Either : 220 V to 240 V<br />

or : 100 V to 127 V<br />

● Insert the plug in the mains socket, the yellow indicator<br />

lights up.<br />

For operation with DC power supply and external battery<br />

(optional)<br />

● Connect optional external battery via cable, see "DC power<br />

operation", page 170.<br />

Note on the use of a socket strip for ancillary equipment<br />

Connecting other equipment to the extension socket strip may<br />

cause the patient leakage current to rise above the permitted<br />

values if a protective earth conductor should fail. The risk of<br />

electric shock cannot be excluded in such cases.<br />

Temporary interruption of power supply<br />

e.g. if hospital reserve power supply is activated.<br />

Without optional Evita DC power supply:<br />

During an interruption of the power supply, <strong>EvitaXL</strong> outputs a<br />

continuous alarm tone for max. 2 minutes. The duration of this<br />

alarm tone may be shorter if <strong>EvitaXL</strong> was switched on for less<br />

than 15 minutes.<br />

<strong>EvitaXL</strong> tolerates power interrupts lasting less than<br />

10 milliseconds – without influencing ventilation.<br />

In the case of power interrupts lasting longer than<br />

10 milliseconds, the machine restarts with a short self-test lasting<br />

about 8 seconds – ventilation is continued with the same<br />

values that were set before the power interruption. If a lower<br />

alarm limit has been set for the minute volume, the »MV low !!!«<br />

alarm is activated until the measured value has risen above the<br />

lower alarm limit.<br />

With optional Evita DC power supply:<br />

See chapter "DC power operation", page 170.<br />

Other equipment, e.g. printers, may only be connected to<br />

the COM port if <strong>EvitaXL</strong> is connected to the mains power<br />

supply via a mains power cable or if it has been earthed<br />

via the earth connection on the back of <strong>EvitaXL</strong>.<br />

Electric power may pose a hazard in all other cases.<br />

30 Instructions for Use <strong>EvitaXL</strong>