Preparing for use<br />
Power supply and connections<br />
Power supply and connections<br />
Electrical power supply<br />
Either : 220 V to 240 V<br />
or : 100 V to 127 V<br />
● Insert the plug in the mains socket, the yellow indicator<br />
lights up.<br />
For operation with DC power supply and external battery<br />
(optional)<br />
● Connect optional external battery via cable, see "DC power<br />
operation", page 170.<br />
Note on the use of a socket strip for ancillary equipment<br />
Connecting other equipment to the extension socket strip may<br />
cause the patient leakage current to rise above the permitted<br />
values if a protective earth conductor should fail. The risk of<br />
electric shock cannot be excluded in such cases.<br />
Temporary interruption of power supply<br />
e.g. if hospital reserve power supply is activated.<br />
Without optional Evita DC power supply:<br />
During an interruption of the power supply, <strong>EvitaXL</strong> outputs a<br />
continuous alarm tone for max. 2 minutes. The duration of this<br />
alarm tone may be shorter if <strong>EvitaXL</strong> was switched on for less<br />
than 15 minutes.<br />
<strong>EvitaXL</strong> tolerates power interrupts lasting less than<br />
10 milliseconds – without influencing ventilation.<br />
In the case of power interrupts lasting longer than<br />
10 milliseconds, the machine restarts with a short self-test lasting<br />
about 8 seconds – ventilation is continued with the same<br />
values that were set before the power interruption. If a lower<br />
alarm limit has been set for the minute volume, the »MV low !!!«<br />
alarm is activated until the measured value has risen above the<br />
lower alarm limit.<br />
With optional Evita DC power supply:<br />
See chapter "DC power operation", page 170.<br />
Other equipment, e.g. printers, may only be connected to<br />
the COM port if <strong>EvitaXL</strong> is connected to the mains power<br />
supply via a mains power cable or if it has been earthed<br />
via the earth connection on the back of <strong>EvitaXL</strong>.<br />
Electric power may pose a hazard in all other cases.<br />
30 Instructions for Use <strong>EvitaXL</strong>