Message Cause Remedy<br />

Fault – Cause – Remedy<br />

O2 supply pressure high !! O2 supply pressure too high. Make sure supply pressure is less than<br />

6 bar.<br />

O2 supply pressure high ! O2 supply pressure too high.<br />

O2 supply is not required for<br />

FiO2 = 21 Vol.%.<br />

Make sure supply pressure is less than<br />

6 bar.<br />

O2 Therapy active ! O2 therapy is activated. Switch off O2 therapy. Exit Standby.<br />

PEEP high !!! Expiratory system obstructed. Check hose system and expiration valve.<br />

Check also for condensate.<br />

Expiratory resistance is increasing. Check bacterial filter.<br />

Replace if necessary.<br />

Machine faulty. Disconnect patient from the device and<br />

continue ventilation without delay using<br />

another independent ventilator.<br />

Call DrägerService.<br />

PEEP valve inop. !!! Internal PEEP valve faulty. Disconnect patient from the device and<br />

continue ventilation without delay using<br />

another independent ventilator.<br />

Call DrägerService.<br />

PPS-Insp. > 4 s !!! The inspiration phase during PPS was<br />

terminated 3 consecutive times after<br />

4 seconds.<br />

Check the ventilation system for leaks.<br />

PPS-Insp. > 1,5 s ! The inspiration phase during PPS was<br />

terminated 3 consecutive times after<br />

1.5 seconds.<br />

Check the ventilation system for leaks.<br />

Pressure limited ! Pmax pressure limit is active. Check the condition of the patient.<br />

Check pattern of ventilation.<br />

Correct setting if necessary.<br />

Pressure meas. inop. !!! Fluid in expiration valve. Replace expiration valve (page 158)<br />

then clean and dry.<br />

Pressure measurement malfunction. Disconnect patient from the device and<br />

continue ventilation immediately with<br />

another independent ventilator.<br />

Call DrägerService.<br />

Procedure overused ? !! Total time of procedures has exceeded<br />

15 minutes within the last hour.<br />

Pulse rate high !!! The pulse rate exceeds the upper alarm<br />

limit.<br />

Pulse rate high ! The pulse rate exceeds the upper alarm<br />

limit.<br />

Pulse rate low !!! The pulse rate has fallen below the lower<br />

alarm limit.<br />

Acknowledge message by pressing<br />

»Alarm Reset« key and confirm. If this<br />

message appears repeatedly: Disconnect<br />

patient from the device and continue<br />

ventilation immediately with another<br />

independent ventilator. Call<br />

DrägerService.<br />

Check the condition of the patient.<br />

Check the ventilation pattern.<br />

If necessary, adjust the alarm limit.<br />

Check the condition of the patient.<br />

Check the ventilation pattern.<br />

If necessary, adjust the alarm limit.<br />

Check the condition of the patient.<br />

Check the ventilation pattern.<br />

If necessary, adjust the alarm limit.<br />