Message Cause Remedy<br />
Fault – Cause – Remedy<br />
O2 supply pressure high !! O2 supply pressure too high. Make sure supply pressure is less than<br />
6 bar.<br />
O2 supply pressure high ! O2 supply pressure too high.<br />
O2 supply is not required for<br />
FiO2 = 21 Vol.%.<br />
Make sure supply pressure is less than<br />
6 bar.<br />
O2 Therapy active ! O2 therapy is activated. Switch off O2 therapy. Exit Standby.<br />
PEEP high !!! Expiratory system obstructed. Check hose system and expiration valve.<br />
Check also for condensate.<br />
Expiratory resistance is increasing. Check bacterial filter.<br />
Replace if necessary.<br />
Machine faulty. Disconnect patient from the device and<br />
continue ventilation without delay using<br />
another independent ventilator.<br />
Call DrägerService.<br />
PEEP valve inop. !!! Internal PEEP valve faulty. Disconnect patient from the device and<br />
continue ventilation without delay using<br />
another independent ventilator.<br />
Call DrägerService.<br />
PPS-Insp. > 4 s !!! The inspiration phase during PPS was<br />
terminated 3 consecutive times after<br />
4 seconds.<br />
Check the ventilation system for leaks.<br />
PPS-Insp. > 1,5 s ! The inspiration phase during PPS was<br />
terminated 3 consecutive times after<br />
1.5 seconds.<br />
Check the ventilation system for leaks.<br />
Pressure limited ! Pmax pressure limit is active. Check the condition of the patient.<br />
Check pattern of ventilation.<br />
Correct setting if necessary.<br />
Pressure meas. inop. !!! Fluid in expiration valve. Replace expiration valve (page 158)<br />
then clean and dry.<br />
Pressure measurement malfunction. Disconnect patient from the device and<br />
continue ventilation immediately with<br />
another independent ventilator.<br />
Call DrägerService.<br />
Procedure overused ? !! Total time of procedures has exceeded<br />
15 minutes within the last hour.<br />
Pulse rate high !!! The pulse rate exceeds the upper alarm<br />
limit.<br />
Pulse rate high ! The pulse rate exceeds the upper alarm<br />
limit.<br />
Pulse rate low !!! The pulse rate has fallen below the lower<br />
alarm limit.<br />
Acknowledge message by pressing<br />
»Alarm Reset« key and confirm. If this<br />
message appears repeatedly: Disconnect<br />
patient from the device and continue<br />
ventilation immediately with another<br />
independent ventilator. Call<br />
DrägerService.<br />
Check the condition of the patient.<br />
Check the ventilation pattern.<br />
If necessary, adjust the alarm limit.<br />
Check the condition of the patient.<br />
Check the ventilation pattern.<br />
If necessary, adjust the alarm limit.<br />
Check the condition of the patient.<br />
Check the ventilation pattern.<br />
If necessary, adjust the alarm limit.<br />