<strong>Living</strong> <strong>Image</strong> ® Software User’s Manual<br />

7.3 Viewing & Exporting Graphical Results<br />

To view a graph of the results:<br />

1. In the Results tab, select an ROI.<br />

2. Click Plot Intensity or Plot Linear Fit.<br />

The linear fit graph plots the logarithm of the intensity, normalized to the selected<br />

source spectrum and the filter transmission properties, against the optical property of<br />

the tissue (μ eff ). The slope of the line is the source depth. If any of the measured points<br />

(in red) deviate significantly from the straight line fit, then the analysis results may be<br />

suspect. The horizontal error bars represent the uncertainty in the optical properties<br />

(usually estimated at ±10%). The vertical error bars represent noise in the image.<br />

The intensity graph displays a graph of the measured intensity in the selected ROI at<br />

each wavelength in the analysis. The intensity is normalized to the selected source<br />

spectrum and the filter transmission properties.<br />

To export graph data:<br />

1. Click the Export Data button .<br />

2. In the dialog box that appears, select a directory for the data and enter a file name<br />

(.csv).<br />

The data can be opened in a spread sheet application such as Microsoft ® Excel ®.<br />