A. IVIS Acquisition Control Panel<br />

Table A.2 Additional IVIS ® System Controls for the IVIS Imaging System 200 Series or IVIS Spectrum<br />

Item Description<br />

Alignment grid Choose this option to activate a laser-generated alignment grid on the stage when the imaging<br />

chamber door is opened. The alignment grid is set to the size of the selected FOV. The grid<br />

automatically turns off after two minutes. If subject alignment is not completed in two minutes,<br />

place a check mark next to Enable Alignment Grid to turn on the grid.<br />

Note: The horizontal cross hair of the alignment grid is offset appropriately to take into account the<br />

height entered in the Subject height box.<br />

Focus Scan Mid <strong>Image</strong> - Choose this option in the Focus drop-down list to set the focal plane at the<br />

maximum dorso-ventral height of the subject at the middle of the animal. This focusing method uses<br />

the laser to scan horizontally across the middle of the subject to determine the maximum subject<br />

height along this line. This option is well suited for animal imaging because the peak height is clearly<br />

identified as the maximum height on the dorsal side along the mid-plane of the animal.<br />

Note: This focusing method is not recommended for microplates or when using a high<br />

magnification field of view (FOV A = 4.0 cm). In these situations, Manual or Subject Size focus<br />

methods are recommended.<br />

Transillumination Setup Choose this option to display the transillumination setup window that enables you to select the<br />

(IVIS Spectrum only)<br />

locations for image acquisition using bottom illumination that originates beneath the stage.<br />

180<br />

Table A.3 Typical field of view (FOV) settings<br />

FOV Setting IVIS Imaging System<br />

Lumina 100 Series<br />

FOV (cm)<br />

200 Series<br />

A 4 10 3.9<br />

B 7 15 6.5<br />

C 10 20 13<br />

D 12 25 19.5<br />