<strong>Living</strong> <strong>Image</strong> ® Software User’s Manual<br />

When efficiency is selected, the fluorescent image data is normalized (divided) by a<br />

stored, calibrated reference image of the excitation light intensity incident on a highly<br />

reflective white plate. The resulting image data is without units, typically in the range of<br />

10 -2 to 10 -9 .<br />

NOTE<br />

F.5 Fluorescent Background<br />

Figure F.7 Illumination profiles for different FOVs on an IVIS Imaging System 100 Series<br />

measured from the center of the FOV.<br />

To enable a more<br />

quantitative<br />

comparison of<br />

fluorescent signals,<br />

choose Efficiency.<br />

Figure F.8 Fluorescent image data displayed in terms of efficiency<br />

On every IVIS ® Imaging System, a reference image of the excitation light intensity is<br />

measured for each excitation filter at every FOV and lamp power. The reference<br />

images are measured and stored in the <strong>Living</strong> <strong>Image</strong> folder prior to instrument<br />

delivery.<br />

Autofluorescence Autofluorescence is a fluorescent signal that originates from substances other than the<br />

fluorophore of interest and is a source of background. Almost every substance emits some<br />

215