[8.B] Administrator options<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 106 of 213<br />

optional settings<br />

The administrator options are only accessible by h/p/cosmos staff <strong>and</strong> authorised service engineers. The access is only<br />

possible with a special code which is not depicted here.<br />

Changing an administrator option might cause serious problems if you are not familiar with all the details. Therefore the<br />

h/p/cosmos service has to be contacted before doing any changes in the administrator options.<br />

For optional settings on devices without UserTerminal an external UserTerminal or a connected PC with the software<br />

h/p/cosmos para control is required.<br />

[8.B1] List of the administrator options<br />

[8.B1] Option Description Comment / Display<br />

OP01 Set back of error messages E 20, E 21,<br />

E 30, E 31, E 32, E 41, E 50 <strong>and</strong> E 51<br />

The required maintenance work has to be performed before deleting<br />

error messages. Confirmation in the display § with "donE"<br />

OP02 Selection of the OEM-Code<br />

This option only sets back the error message. If the error still exists,<br />

the error message will appear again<br />

By selecting the OEM code the functions <strong>and</strong> settings of the<br />

UserTerminal will be adapted to the requirements of the listed OEM<br />

partner.<br />

Following OEM codes can be selected:<br />

0 = h/p/cosmos (St<strong>and</strong>ard)<br />

1 = JAEGER / VIASYS / CARDINAL HEALTH<br />

2 = Ergo-Fit (1997-98)<br />

3 = Proxomed / Kardiomed<br />

4 = KISTLER Gaitway<br />

6 = COSMED<br />

9 = SCHILLER<br />

OP03 Selection of device type<br />

Not all companies listed above are or were OEM partners of<br />

h/p/cosmos.<br />

Selection of the correct treadmill device type according to the „list of<br />

device types” in the appendix. On the ladder-ergometer only the device<br />

type 0.1 <strong>and</strong> 1.1 can be confirmed..<br />

Note: Selection <strong>and</strong> confirmation of a device type sets back all user<strong>and</strong><br />

administrator options to default.