file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 109 of 213<br />

optional settings<br />

[8.B1] OP13 Activating <strong>and</strong> de-activating motor break The motor break prevents the <strong>running</strong> belt from manual moving (for<br />

example because of the subjects bodyweight) when the speed value is<br />

set to “0”. If the treadmill recognizes a <strong>running</strong> belt movement it<br />

activates the motor break for 10 seconds.<br />

.<br />

OP14 Maximum acceleration time for RS232<br />

(option only available for treadmills, not<br />

for ladder ergometers)<br />

OP16 Activating <strong>and</strong> de-activating heart rate<br />

measurement<br />

Remark:<br />

The motor break is only working at <strong>running</strong> <strong>machine</strong>s with integrated<br />

speed measurement. The speed sensor must be switched on<br />

(OP11=1). If the <strong>running</strong> <strong>machine</strong> is not equipped with speed<br />

measuring system this could be retrofitted easily.<br />

To activate the motor break it is necessary to additionally change<br />

some parameters at the frequency inverter. Ask the service<br />

department of h/p/cosmos sports & medical gmbh for detailed<br />

instruction.<br />

Attention!<br />

If the device is switched off or disconnected from<br />

the power supply the motor break is not working!<br />

After 10 seconds of operation of the motor break,<br />

the break is released automatically <strong>and</strong> is only<br />

activated again after a belt movement. A short jerk<br />

may occure during reactivation.<br />

§ shows OFF = Motor break not active<br />

§ shows ON = Motor break active<br />

§ shows: e.g. 5 (max. acceleration-increment)<br />

$ shows: e.g. 131.1 (max. acceleration-time in sec.)<br />

% shows: SEC.<br />

= reports ...<br />

0 = heart rate measurement is de-activated / OFF<br />

1 = heart rate measurement is activated, circuit board HFU1 or<br />

RE06REC (default)<br />

2 = heart rate measurement is activated, circuit board HFUi<br />

auto = automatic detection <strong>and</strong> activation of circuit board<br />

4 = Polar W:I:N:D (further settings in user OP19 <strong>and</strong> OP23 necessary)<br />

OP33 Distance between rungs of the ladder For correct calculation of the number of rungs / steps, the distance<br />

ergometer<br />

between two rungs can be changed <strong>and</strong> confirmed in this option.<br />

(option only available for ladder St<strong>and</strong>ard: 254 for 25,4 cm<br />

ergometers, not for treadmills)