installation & commissioning<br />

Immediately at first installation an electric safety test / measurment<br />

has to be performed for „Protective Earth Resistance“, „Isolation<br />

Resistance“ <strong>and</strong> „Leakage Current“ <strong>and</strong> the values have to be<br />

recorded on a special protocol [cos11690xx] <strong>and</strong> marked with „first<br />

measured values“. For details please refer to chapter<br />

„maintenance / inspections“. One copy of this protocol<br />

[cos11690xx] remains at the owner’s manual <strong>and</strong> the original with<br />

the „first measured values“ shall be sent to h/p/cosmos. The<br />

potential equalization cable has to be<br />

accomplished �<br />

connected AFTER the measurements.<br />

Lubricate the <strong>running</strong> <strong>machine</strong> before introduction according to<br />

instruction with 30 ml special silicon oil which is part of the “service<br />

box”.<br />

Check the tension of the <strong>running</strong> belt (no slip).<br />

Adjust the <strong>running</strong> belt properly before instruction with the<br />

customer.<br />

The allen-key must be pulled out of the screw<br />

immediately after usage due to risk of injury!<br />

Test the function of the heart rate transmission with the help of a<br />

simulator or a POLAR sender. Search for interferences <strong>and</strong><br />

possible reasons.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 60 of 213<br />

accomplished �<br />

accomplished �<br />

accomplished �<br />

accomplished �

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!