14.11.2012 Views

running machine - H-P-COSMOS Sports and Medical

running machine - H-P-COSMOS Sports and Medical

running machine - H-P-COSMOS Sports and Medical

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 110 of 213<br />

optional settings<br />

[8.B1] OP34 Manual speed calibration<br />

Normally the speed calibration values are set by selecting the correct<br />

(counted increments from the speed device type (administrator option number 3). Only for special speed<br />

sensor for 10 m belt movement) <strong>and</strong> special models the speed calibration values have to be changed. If<br />

you change the speed calibration value, please also pay attention to<br />

the correct settings in option number 5 (firmware speed limit), number<br />

11 (speed measurement; must be switched on) <strong>and</strong> option number 48<br />

(Frequency inverter speed limit) of the administrator mode. To change<br />

the speed calibration value please use the + <strong>and</strong> – buttons. For quick<br />

movements between st<strong>and</strong>ard values, please use the UP- <strong>and</strong> DOWNbuttons.<br />

If the speed measurement is switched off in OP11, then OP34 will only<br />

report internal program absolutes. In the other case following values<br />

are stored:<br />

N Running <strong>machine</strong> series h/p/cosmos mercury with<br />

special speed 10 km/h:<br />

Speed Calibration Value: 8.490<br />

Inverter Frequency: 80.2 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos mercury with<br />

speed 22 km/h:<br />

Speed Calibration Value: 5.988<br />

Inverter Frequency: 123.3 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos mercury with<br />

special speed 30 km/h:<br />

Speed Calibration Value: 4.165<br />

Inverter Frequency: 119.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 10 km/h:<br />

Speed Calibration Value: 7.800<br />

Inverter Frequency: 80.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 22 km/h:<br />

Speed Calibration Value: 4.495<br />

Inverter Frequency: 94.5 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

speed 25 km/h:<br />

Speed Calibration Value: 4.495<br />

Inverter Frequency: 110.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 30 km/h:<br />

Speed Calibration Value: 4.010<br />

Inverter Frequency: 115.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 40 km/h:<br />

Speed Calibration Value: 2.874<br />

Inverter Frequency:<br />

Continuation next page…<br />

113 Hz