[12.S] Digital (serial) RS232 interface / USB converter<br />
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />
© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 185 of 213<br />
accessories & options<br />
Connecting a <strong>running</strong> <strong>machine</strong> to a medical device results in a medical system. Always use IEC<br />
601-1 approved potential isolation components when linking medical devices via interface. Only<br />
authorised trained staff is entitled to perform this connection. This medical system, when<br />
connecting a <strong>running</strong> <strong>machine</strong> to a medical device, has to be connected via a potential-balance-<br />
cable with the provided connector bolt <strong>and</strong> with the provided bearing within the medically used<br />
room.<br />
The interface serves as a data transmitter from the <strong>running</strong> <strong>machine</strong> to a PC or other controlling devices. You can also<br />
control the Running-Machine from another controlling device (ECG. o.a.) or with the appropriate Software from a PC, via<br />
interface. St<strong>and</strong>ard for all devices is ONE serial interface COM 1. Optionally there are further serial interfaces COM 2,<br />
COM3 <strong>and</strong> COM4 available. For further information please contact your local dealer or h/p/cosmos. The interface<br />
protocol h/p/cosmos coscom © is available on request or can be found at www.coscom.org. The incorporated protocols<br />
(e.g. for ECG-devices) are listed in the list of options under OPTION 20 / OPTION 21. See the compatibility list of the<br />
RS232 interface in the chapter „Technical data“. More recent computers will have only an USB interface instead of the<br />
RS232 interface. For that case an „USB to RS232 interface-adapter-cable“ is available at h/p/cosmos under order<br />
number cos12769. To control via USB interface the processor must be a Pentium 1.8 GHz or higher.<br />
[12.S1] Interface cable RS232 for connection to PC<br />
To be used for h/p/cosmos para graphics ® , h/p/cosmos para control ® , PC, ECGs, Ergospirometry devices. For<br />
connecting your PC to the <strong>running</strong> <strong>machine</strong> you need an interface cable (optional equipment) 9-pole Sub- D (male<br />
- female) with crossed transmission- receive- cable (PIN 2 <strong>and</strong> PIN 3).<br />
6<br />
7<br />
8<br />
9<br />
1<br />
2<br />
3<br />
4<br />
5<br />
PC COM1 oder COM2<br />
CONNECTOR DB9<br />
Buchse / female<br />
[12.S2] Interface cable RS232 for connection to ECG Schiller<br />
R x D<br />
T x D<br />
GND<br />
[cos00097010034] for 5 m / [cos00097010035] for 10 m<br />
1<br />
2<br />
3<br />
4<br />
5<br />
6<br />
7<br />
8<br />
9<br />
Laufb<strong>and</strong> / treadmill<br />
CONNECTOR DB9<br />
Stecker / male<br />
Connection cable between <strong>running</strong> <strong>machine</strong> <strong>and</strong> ECG Schiller AT 10 / AT 60 or CS 100 / CS 200<br />
6<br />
7<br />
8<br />
9<br />
1<br />
2<br />
3<br />
4<br />
5<br />
EKG / ECG Schiller<br />
CONNECTOR DB9<br />
Stecker / male<br />
R x D<br />
T x D<br />
GND<br />
[cos00097010036] for 5 m<br />
1<br />
2<br />
3<br />
4<br />
5<br />
6<br />
7<br />
8<br />
9<br />
Laufb<strong>and</strong> / treadmill<br />
CONNECTOR DB9<br />
Stecker / male