[12.L] Safety arch with chest belt <strong>and</strong> harness<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 178 of 213<br />

accessories & options<br />

N Every time before using the safety arch with harness a visual examination of the system has<br />

to done (examination for damages <strong>and</strong> wear). Especially the rope, the harness <strong>and</strong> all links<br />

like the snap hook <strong>and</strong> the rope brake have to be checked.<br />

N All wearing parts of the system (rope, harness <strong>and</strong> all links like the snap hook <strong>and</strong> the rope<br />

brake) have to be changed immediately in case of any damages or wear. At least every two<br />

years all wearing parts have to be changed.<br />

Indispensable accessory for stress testing, accident<br />

prevention <strong>and</strong> for the general safety of athletes: The<br />

safety arch with chest belt <strong>and</strong> harness. Automatic stop of<br />

the <strong>running</strong> belt in the moment a fall occurs. The chest<br />

belt secures the subject <strong>and</strong> prevents falling over to the<br />

front. The h/p/cosmos safety arch is available for all<br />

h/p/cosmos <strong>running</strong> <strong>machine</strong>s.<br />

The chest belt is connected via harness with the<br />

emergency stop at the crossbar of the safety arch. If<br />

there is more than 8 kg / 17.6 lb traction pulling on the<br />

harness, the switch will stop the <strong>running</strong> belt immediately.<br />

The length of the harness is individually adjustable.<br />

The chest belt straps do not pinch or restrict the athlete. The chestbelt has to be<br />

put on with the h/p/cosmos logo at the front. The vertical belt (1), to attach the<br />

carabiner, has to be put on below the shoulder<br />

belts (2), facing the body, to unload the seem at<br />

the joint (3). If the chestbelt is put on back to front<br />

or inside out or twisted, a shearing strain can occur<br />

at the joints <strong>and</strong> then lead to a loosening of the<br />

joints.<br />

Despite the fast stop of the <strong>running</strong> belt skin-abrasion at the legs might occur due<br />

to the <strong>running</strong> out of the belt, because a fallen test person still touches <strong>running</strong> out<br />

<strong>running</strong> belt for a few seconds.<br />

Nevertheless the risk of injuries is minimized drastically compared to a fall at systems without the arch <strong>and</strong> compared<br />

with falls outdoors. The maximum body weight is 140 kg / 309 lbs.<br />

The ceiling height should be at least 2.46m / 97.6 inches (0 % elevation) rsp. 2.57 / 101.6 inches (24 %) if used with<br />

h/p/cosmos mercury <strong>and</strong> 2.51 / 96.5 inches (0 %) rsp. 2.68 m / 105.5 inches (max. elevation) if used with <strong>running</strong><br />

<strong>machine</strong>s h/p/cosmos quasar or h/p/cosmos pulsar.