[12.Q] Monitor / laptop arm<br />

UserTerminal of<br />

oversize model<br />

families<br />

200-450/75-300<br />

(venus/saturn).<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

A movable monitor <strong>and</strong> laptop arm for the external control unit<br />

of the oversize <strong>running</strong> <strong>machine</strong>s h/p/cosmos saturn <strong>and</strong><br />

h/p/cosmos venus is optionally available. Using this arm, the<br />

monitor also can be turned in the field of vision of the athlete or<br />

patient.<br />

Monitor arm (movable) for PC monitor or Laptop<br />

[cos13320-01]<br />

Laptop computer [cos13476]<br />

[12.R] Safety multiple socket<br />

The stable safety multiple socket [cos100157] can be used for<br />

the safe connection of a medical h/p/cosmos treadmill with<br />

peripheral devices such as PCs, printer or monitors.<br />

It meets all requirements for multiple sockets for creating<br />

connections of a medical electrical system within the patients<br />

environment.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 184 of 213<br />

accessories & options<br />

safety multiple socket [cos100157]<br />

As claimed in the norm EN60601-1 for medical electrical devices, the multiple socked is accessible only with the help of<br />

a tool. With this method an unintentionally made connection of further devices to the system is avoided, which lowers<br />

the risk of an exceedance of the maximum allowed patient-leakage-current of 0.5mA at the complete system.<br />

N Due to the very high load of treadmills it is never allowed to connect more than 1 treadmill or<br />

other consumers with high load do the safety multiple socket. Use a dedicated line only.<br />

N The power consumption of the consumers which are connected in addition to the treadmill-<br />

ergometer to the safety multiple socket must be in total lower than 300W.<br />

N When connecting treadmills to the safety multiple socket, there are no performance-limitations<br />

to expect at low-performance applications (walking, slow <strong>running</strong>). For high-performance<br />

applications (fast <strong>running</strong>, sprints, etc.) the use of the safety multiple socket can result in<br />

performance limitations.<br />

N For high performance applications a 3-phase powered treadmill is recommended.