[9.G2] Driving belt with timing-belt system<br />

maintenance & safety inspections<br />

Timing belt drives have no automatic belt adjuster. The tension of the timing belt (depending on the model) needs<br />

to be adjusted by the manual belt adjuster or by changing the position of the drive motor.<br />

The driving belt tension can be controlled according to point control of <strong>running</strong> belt tension:<br />

N Block the <strong>running</strong> belt at speed of 1 to 1.5 km/h.<br />

N No backlash must be between the motor shaft <strong>and</strong> the driving belt.<br />

N Check if it is possible to move the drive belt with thumb <strong>and</strong> forefinger 90° sidewards by using moderate<br />

force. If the angle is more or less than 90°, the tension is too high or too low <strong>and</strong> the drive belt must be<br />

adjusted accordingly.<br />

N The noise (at higher speeds) <strong>and</strong> the straight position of the driving belt need to be considered.<br />

[9.H] Side step platforms: Check for non-slipping<br />

Each <strong>running</strong> <strong>machine</strong> has an anti-skid area on both sides next to the <strong>running</strong> surface. These step platforms offer safe<br />

halt when leaving the <strong>running</strong> surface in emergency cases. Check the anti-skid areas in regular intervals <strong>and</strong> replace<br />

them at signs of wearing off.<br />

[9.I] Hygiene, cleansing, cleansing of the interior<br />

For safety reasons switch off the device <strong>and</strong> pull the plug before cleansing, disinfecting <strong>and</strong><br />

before opening it. Pay attention to the instruction manual of the device.<br />

Pay attention to the notes of the operating instructions of the concerning devices <strong>and</strong> the accessories.<br />

Chemicals needed for application or cleansing have to be stored in appropriate reservoirs because of the danger of<br />

mixing them up. Interior Cleansing is required in intervals of 6 months. Depending on your model lift up the motor hood<br />

by lifting it up or remove bellow from upper frame.<br />

Clean the interior of the <strong>running</strong> <strong>machine</strong> by removing the dirt with the help of a vacuum cleaner. Pay special attention to<br />

the ventilation net of the drive motor <strong>and</strong> the cover of the frequency inverter. At devices with external control unit the<br />

internal space of the control unit must be cleaned as well.<br />

The cleaning of the surfaces of the treadmill can be done with a light moistened cloth. For disinfection h/p/cosmos<br />

recommends Bacillol plus [cos12179]. Before using cleaning <strong>and</strong> disinfection substances always test the compatibility<br />

of the material at bad viewable places.<br />

In case patients with communicable diseases are treated on the treadmill (or there is only the suspicion for<br />

communicable diseases), the treadmill has to be disinfected at the disrcretion of the doctor or medical stuff before <strong>and</strong><br />

after ther treatment.<br />

This covers all parts the patient might have contact to, such as front- <strong>and</strong> side-h<strong>and</strong>rails, arm support, user terminal<br />

(keyboard <strong>and</strong> display), emergency stop, chestbelt or other heart rate measurement systems, <strong>running</strong> belt, emergency<br />

stop with magnet switch, safety harness for fall stop prevention, vest for unweighting system, motor cover, frameparts,<br />

treads, wheelchair ramp, drinking bottle, etc.<br />

Cleaning procedure:<br />

Spray contact surfaces, impact, rub off.<br />

Spray disinfectant on a tissue <strong>and</strong> rub clean the possibly contaminated area.<br />

Always follow the disinfectants manufacturer’s manual.<br />