[8.B2] St<strong>and</strong>ard settings of administrator options<br />
[8.B2] Option Description St<strong>and</strong>ard setting Adjustable Range<br />
OP 01 Set back of error messages E 20, E 21,<br />
E 30, E 31, E 32, E 41, E 50 <strong>and</strong> E 51<br />
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />
© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 119 of 213<br />
optional settings<br />
OP 02 OEM-Code depending on the model 0 = h/p/cosmos (St<strong>and</strong>ard)<br />
1 = JAEGER / VIASYS /<br />
Cardinal Health<br />
2 = Ergo-Fit (1997-98)<br />
3 = Proxomed / Kardiomed<br />
4 = KISTLER Gaitway<br />
6 = COSMED<br />
9 = SCHILLER<br />
OP 03 Selection of device type<br />
- see list [6.B3] -<br />
depending on the model (depending on the model)<br />
OP 04 Maximum speed for reverse belt rotation depending on the device type in<br />
OP03<br />
OP 05 Maximum Speed<br />
depending on the device type in<br />
OP03<br />
1.0 km/h .... max. speed<br />
1.0 km/h .... max. speed<br />
OP06 Max. acceleration time in seconds Increment 131 = 5 sec. Increments: 255 ... 1<br />
Seconds: 2.6 ... 655.4<br />
OP 07 Maximum Elevation<br />
depending on the device type in<br />
OP03<br />
0 .... max. elevation<br />
OP 08 Heart Rate Display Interval 0 = beat to beat 0 or 1 ... 9 seconds<br />
OP 09 Display Test for all displays<br />
OP 10 Enter the serial number of the device 300… 12-digit serial number<br />
OP 11 Activating <strong>and</strong> de-activating speed depending on the device type in 0 = without speed measurement<br />
measurement<br />
OP03<br />
1 = with speed measurement<br />
OP 12 Setting of allowance for speed-signal- 6 %<br />
0 = no comparison<br />
comparison<br />
This value does not reflect the 6 = 6 % deviation allowed<br />
(Measurement / Control)<br />
speed accuracy of the <strong>running</strong>- 8 = 8 % deviation allowed<br />
belt.<br />
10 = 10 % deviation allowed<br />
OP13 Motor brake OFF OFF = not active<br />
ON = active<br />
OP 14 Maximum acceleration time for RS232 Increment 131 = 5 sec Increments: 255 ... 1<br />
Seconds: 2.6 ... 655.4<br />
OP 16 Activating <strong>and</strong> de-activating heart rate (depending on the model) 0 = HF measurement de-activated<br />
measurement<br />
1 = HF measurement activated,<br />
HFU1 + RE06REC<br />
2 = HF measurement activated,<br />
HFUi;<br />
auto = autom. detection<br />
OP 33 Distance between rungs (only for ladder<br />
ergometer)<br />
254 = 25.4 cm Distance between rungs in 0.1 cm<br />
OP 34 Manual Speed Calibration<br />
(depending on the device type;<br />
OP 03)<br />
Range from: 0 to 65535