accessories & options<br />
The <strong>running</strong> belt rotation reverse will be announced by an acoustic signal <strong>and</strong> the display = "r." Speed will be<br />
indicated as a negative value while in operation = „-0.0 ... -5.0“.<br />
[12.M1] Pull Cord Actuator (lanyard + magnet)<br />
This stop function is a safety feature that is<br />
used with a waist belt. Especially with<br />
reverse belt rotation (downhill) it is also<br />
important to connect. The athlete or patient<br />
is connected to the magnetic switch at the<br />
UserTerminal via a ripcord at the waist belt.<br />
In case of an emergency the magnet gets<br />
ripped of from the UserTerminal <strong>and</strong> the<br />
<strong>running</strong> <strong>machine</strong> stops immediately. The<br />
ripcord is not a prevention of falling. At<br />
removed magnet „PULL STOP“ will occur in<br />
the display.<br />
Safety harnesses such as safety stop cord type (safety lanyard with pull-cord actuator) <strong>and</strong>/or safety arch with<br />
harness, chest belt <strong>and</strong> stop function <strong>and</strong>/or unweighting systems with patient vest <strong>and</strong> harness shall be adjusted<br />
in a way, that the subject can not use the rear 30% of the <strong>running</strong> deck length.<br />
Pull Cord Actuator (lanyard + magnet) with STOP function: [cos00098050043]<br />
Chest-harnesses, patient-vests, waist-belts with leash <strong>and</strong> cuffs are not designed for direct<br />
skin or mucous membrane contact. Wear appropriate clothing.<br />
picture type of stop function interface<br />
communication<br />
Pull-cord<br />
actuator (lanyard<br />
with<br />
magnetic switch<br />
on the<br />
UserTerminal <strong>and</strong><br />
waist belt on the<br />
subject)<br />
N magnet is pulled off through<br />
the cord<br />
N <strong>running</strong> belt & inclination<br />
system stop fast<br />
N <strong>machine</strong> is NOT cut off from<br />
main power of drive motor<br />
<strong>and</strong> inclination system<br />
N display active <strong>and</strong> shows<br />
“PULL STOP”<br />
N interface active, but no start,<br />
speed or elevation signal<br />
executed<br />
N keyboard inactive<br />
communication<br />
NOT interrupted<br />
during<br />
“pull stop” stage<br />
how to restart<br />
N reconnect magnet<br />
to its position on<br />
UserTerminal<br />
N press START key,<br />
N or new comm<strong>and</strong><br />
through interface