14.11.2012 Views

running machine - H-P-COSMOS Sports and Medical

running machine - H-P-COSMOS Sports and Medical

running machine - H-P-COSMOS Sports and Medical

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 108 of 213<br />

optional settings<br />

[8.B1] OP08 Heart Rate Display Interval § indicates: Heart Rate Display Interval in seconds<br />

Einstellbereich: 0 (beat to beat) .... 1 bis 9 Sekunden<br />

Adjustable from 0 (beat to beat) .... to 9 seconds<br />

Compare: An original POLAR watch has a display interval of 5<br />

seconds<br />

$ reports: SEC<br />

OP09 Test of all displays <strong>and</strong> LEDs<br />

OP10 Enter the serial number of the device Change the values with a or d<br />

according to the nameplate<br />

Confirm the values with c<br />

OP11 Activating <strong>and</strong> de-activating speed<br />

Example: Treadmill with serial number cos30000va08-0128 on the<br />

nameplate<br />

§ reports: 300 (item number of device)<br />

$ indicates: 0008 (item number of device)<br />

% indicates: 0128 (consecutive serial numbers of this device<br />

family)<br />

Normally the speed measurement is activated / de-activated by<br />

measurement<br />

selecting the correct device type (administrator option number 3). Only<br />

in very special cases a special adjustment in option number 11 is<br />

necessary. Speed measurement can only be activated if the device<br />

has a speed sensor<br />

§ reports:<br />

0 = without speed measurement / -sensor<br />

1 = with speed measurement / -sensor<br />

(permanent control)<br />

2 = with speed measurement / -sensor<br />

(control as of speed > 2 km/h, no control < 2km/h)<br />

OP12 Setting of allowance for speed-signal- The MCU5 control unit is comparing continuously the data of the<br />

comparison (Measurement / Control) speed measurement with a calculated reference value. In order to<br />

detect problems with the speed measurement or speed control or e.g.<br />

These values do not reflect the real performance limitations due to a bad voltage supply, an allowance<br />

speed accuracy of the <strong>running</strong> belt. range can be set in this option. In case the allowance is exceeded the<br />

Accuracy of <strong>running</strong>-belt speed is higher. error message E.31 (measured speed is too high) or E. 32 (measured<br />

speed is too low) is reported.<br />

§ reports: OFF = no comparison<br />

6 = 6 % deviation is allowed<br />

8 = 8 % deviation is allowed<br />

10 = 10 % deviation is allowed