[9.D2] Lubricant test<br />
maintenance & safety inspections<br />
The condition of the oil film between <strong>running</strong> surface <strong>and</strong> <strong>running</strong> belt can be established with the lubricant-test<br />
[cos14379] which is available at h/p/cosmos. The lubricant-test consists of a pre-wheighed commercial paper<br />
tissue. The tissue needs to be pulled through the gap between <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface in a defined lenght<br />
in order to absorb a certain quantity of lubricant. Based on previous experience h/p/cosmos is able to evaluate the<br />
quantity of the oil film by measuring the weight difference. However, it needs to be considered that this test can<br />
only find out approximate values <strong>and</strong> trends <strong>and</strong> can never replace the direct survey by the authorized h/p/cosmos<br />
technician in the prescribet time intervals.<br />
For security reasons, the <strong>running</strong> <strong>machine</strong> needs to be switched off <strong>and</strong> the power plug needs<br />
to be pulled out for full separation from the mains.<br />
[9.D2] Step Action Reaction/display<br />
[01]<br />
[02]<br />
Trittfläche / footboard<br />
1 meter<br />
Laufgurt / <strong>running</strong> belt<br />
Trittfläche / footboard<br />
L<br />
Taschentuch<br />
Tissue<br />
M<br />
R<br />
10 cm<br />
10 cm<br />
There are 3 pre-weighed paper tissues<br />
[cos14379] required for the test.<br />
Every single tissue is located in a separate<br />
plastic bag. The information (blue letters) in<br />
the left picture need to be entered to the<br />
sticker which is on the plastic bag.<br />
The tissue-test needs to be performed at 3<br />
places under the <strong>running</strong> belt:<br />
- Left edge<br />
- center<br />
- right edge<br />
For each test a separate tissue needs to be<br />
used!<br />