14.11.2012 Views

running machine - H-P-COSMOS Sports and Medical

running machine - H-P-COSMOS Sports and Medical

running machine - H-P-COSMOS Sports and Medical

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

maintenance & safety inspections<br />

N Checking the tightening of the drive belt with the „frequency-test“<br />

Checking with measuring device for belt tension [cos14863]:<br />

N Switch off the <strong>running</strong> <strong>machine</strong> at the main switch.<br />

upper trumm<br />

N Visual inspection of the drive belt for abraison <strong>and</strong><br />

correct fitting.<br />

N Switch on the tuning device.<br />

N The tuning device must be in mode „BASS“,<br />

change with button GUITAR/BASS.<br />

N Hold the tuning device with the microphone (MIC)<br />

close to the bottom trumm of the drive belt <strong>and</strong><br />

pick the trumm (see picture).<br />

N The tension of the drive belt at the bottom trumm<br />

(see picture) needs to be between 98 Hz <strong>and</strong><br />

103 Hz (frequency during swinging).<br />

bottom trumm / place to pick<br />

N Check the tone in the display: Target = 1G – adjust the tension of the drive belt if necessary.<br />

N Check the amplitude of the needle: Target 440 ... 460 – adjust the tension of the drive belt if necessary.<br />

N When the tuning device is indicating the tone 1G <strong>and</strong> an amplitude of the needle of 440 ... 460 the tension of<br />

the drive belt is ok.<br />

Chart with tones to the according frequency:<br />

target value frequency drive belt tone<br />

LB<br />

scale<br />

70 2D 420<br />

76 2D 455<br />

79 4E 425<br />

81 4E 440<br />

82 4E 440<br />

85 4E 447<br />

87 4E 457<br />

90 1G 422<br />

92 1G 425<br />

95 1G 432<br />

X 98 1G 440<br />

X 100 1G 450<br />

X 103 1G 457<br />

104 3A 425<br />

106 3A 427<br />

109 3A<br />

HC<br />

437<br />

The target value of the frequency is between 98Hz <strong>and</strong> 102Hz. This is equal to the tone 1G. If the result of the test<br />

is out of this area, the tension of the drive belt must be adjusted.<br />

frequency < value 98: belt tension need correction.<br />

frequency > value 102: belt tension need correction.<br />

After changing the tension of the drive belt new check is necessary.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 137 of 213