[11.B] Running <strong>machine</strong>s, 170-190/65<br />
UserTerminal<br />
(Display & Keyboard)<br />
Power supply<br />
(st<strong>and</strong>ard, read nameplate)<br />
Dedicated line required.<br />
h/p/cosmos stellar<br />
h/p/cosmos stellar lt<br />
h/p/cosmos quasar<br />
h/p/cosmos quasar lt<br />
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />
© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 156 of 213<br />
h/p/cosmos stellar med<br />
h/p/cosmos stellar med lt<br />
6 LCD displays / 6 keys / display of mode <strong>and</strong> unit with LED.<br />
„lt-models“ have no UserTerminal (no display, no keyboard). Control only via interface.<br />
h/p/cosmos quasar med<br />
230 Volt AC 1~/N/PE f: 50/60 Hz 3680 VA<br />
Fuse (st<strong>and</strong>ard, read nameplate) 16 A<br />
Power consumption (St<strong>and</strong>ard,<br />
read nameplate)<br />
15.0 A 15.5 A 15.0 A 15.5 A 16 A 8.0 A<br />
h/p/cosmos quasar med lt<br />
h/p/cosmos pulsar<br />
h/p/cosmos pulsar lt<br />
technical data<br />
h/p/cosmos pulsar 3p<br />
h/p/cosmos pulsar 3p lt<br />
400 Volt AC3~<br />
/N/PE f:50/60Hz<br />
11000 VA<br />
Power input 3450 VA 3565 VA 3450 VA 3565 VA 3680 VA 5536 VA<br />
Drive motor capacity 3300 W 4300 W<br />
Elevation motor capacity not applicable 500 W not applicable 500 W 500 W<br />
Safety st<strong>and</strong>ards c<br />
IEC EN 60335-1(VDE 0700)<br />
IEC EN 60601-1-2,<br />
IEC EN 60601-1-4, VDE 0701,<br />
EN 957-1, EN 957-6<br />
n<br />
MDD, directive 93/42/EEC + 2007/47/EC,<br />
IEC EN 60601-1 (VDE 0750), VDE 0751,<br />
IEC EN 60601-1-2, IEC EN 60601-1-4,<br />
EN 957-1, EN 957-6<br />
Leakage current 0.6 mA 0.2 mA 0.2 mA 0.25 mA<br />
Isolation transformer --- 2000 VA 2000 VA 3 x 1500 VA<br />
Degree of protection class I M / --- / IP 20 class I M / type B R / IP 20<br />
Overvoltage category Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt<br />
Mode of operation acc. IEC 601-1 continuous operation<br />
Field of application / Accuracy sports <strong>and</strong> fitness: SA sports <strong>and</strong> medical: SIA<br />
Classification according to MDD --- --- II b<br />
Speed<br />
Definition<br />
0.0 ... 25.0 km/h<br />
0.1 km/h<br />
0.0 ... 25.0 km/h<br />
0.1 km/h<br />
Allowance* +/- 5 %<br />
0.0 ... 40.0 km/h<br />
0.1 km/h<br />
Option special speed 0.0 ... 10.0 km/h , 0.0 ... 40.0 km/h , 0.0... 45.0 km/h 0.0 ... 10.0 km/h or 0.0... 45.0 km/h<br />
Angle of inclination**<br />
Definition<br />
not applicable 0.0 ... 28.0 %<br />
0.1 %<br />
not applicable 0.0 ... 28.0 %<br />
0.1 %<br />
- 25.0 % ... + 25.0 %<br />
0.1 %