safety, warnings, precautions, prohibitions<br />

Examples of ME equipment <strong>and</strong> ME systems on the network <strong>and</strong> data network:<br />

• modality connectivity via network to PACS <strong>and</strong> WEB-based electronic image distribution<br />

• server-based database for long-term ECG devices or stress-test systems<br />

• telemedicine applications<br />

• Wi-Fi connectivity of patient monitors<br />

• et.al.<br />

Appendix A of DIN EN 14971 includes questions to identify characteristics of a medical device that may have security<br />

implications. The application of this Annex on the causes <strong>and</strong> risks of networks <strong>and</strong> data network should at least,<br />

according to DIN EN 60601-1, include the following considerations or risk <strong>and</strong> potential causes respectively:<br />

• Telemedicine <strong>and</strong> tele-service with external access to internal network or data network of an operator (hospital<br />

or other)<br />

• Remote service of modalities manufacturers<br />

• compatibility of operating systems<br />

• changes <strong>and</strong> upgrades of the software (operating system, applications, etc.)<br />

• The impact <strong>and</strong> consequences of patch management<br />

• Interface management (e.g. intolerance of 10 MB network cards with a 100 MB Ethernet network or other)<br />

• Connections (modification of hardware, network connector)<br />

• protocols such as DICOM, HL7 in the network or data network<br />

• Packet address structure <strong>and</strong> b<strong>and</strong>width<br />

• Heterogeneous network topology<br />

• Normal network traffic <strong>and</strong> b<strong>and</strong>width requirements<br />

• peak network load<br />

• security <strong>and</strong> long-term readability of data carriers<br />

• Security against destructive software, unauthorized software updates or upgrades<br />

• Maximum acceptable response time<br />

• Acceptable error rate of the network or data network<br />

• availability during planned <strong>and</strong> unplanned maintenance<br />

• inconsistency of interfaces <strong>and</strong> formats, which can lead to loss of accuracy during the data transfer<br />

• etc.<br />

Annex D of the DIN EN 14971 describes examples of risks <strong>and</strong> other factors in connection between the ME device <strong>and</strong><br />

network or data network:<br />

• What predictable abuse can occur?<br />

• Is the connection to the network or data network carried out in accordance to the intended use or purpose<br />

according to § 3 para 10 of the <strong>Medical</strong> Devices Act, or is it in contradiction?<br />

• Could a false data flow occur from or to any connected PEMS?<br />

• What shall the medical data transferred through the network or data network achieve or what shall happen to<br />

them? What happens if the network or data network collapses during data transmission?<br />

• Can there be any deviations from specified operating characteristics of each participating PEMS?<br />

• What qualities <strong>and</strong> operating characteristics does a PEMS have <strong>and</strong> how <strong>and</strong> in what form may they be<br />

influenced by the network or data network?<br />

• Does a complete description of the parameters of the network or data network exist, such as network topology,<br />

configuration, parameters, b<strong>and</strong>width (100 MB Ethernet, 1GB Ethernet, etc.), etc.?<br />

• Can an overloading of the network or data connections within the network node occur?<br />