Medium to achieve hygienic safety:<br />

Bacillol plus (h/p/cosmos order number cos12179)<br />

> See manufacturers safety warnings <strong>and</strong> data sheet<br />

Depollution of used chleaning medium:<br />

Bacillol plus (h/p/cosmos order number cos12179)<br />

> Manufacturers depollution regulations<br />

[9.J] Cleansing <strong>and</strong> adjustment of the speed-sensor/Light-barrier<br />

maintenance & safety inspections<br />

Depending on the model, the firmware version <strong>and</strong> the manufacturing date, the <strong>machine</strong>s are equipped with …<br />

N no speed sensor (so no cleansing required)<br />

N Light barrier <strong>and</strong> incremental waver mounted without housing at the front (drive shaft) of the drive engine.<br />

N Device from the construction year app. 2003 onwards have instead of the light barriers an induction sensor with<br />

metal toothed lock washer.<br />

N Only at oversize <strong>running</strong> <strong>machine</strong>s with a <strong>running</strong> surface of 200/75cm up to 300/125cm <strong>and</strong> up to the<br />

construction year 2002: Speed sensor mounted in its own housing at the rear part (drive shaft) of the drive engine.<br />

Due to the aspect, that the sensors have their own housing, they must not be cleaned.<br />

The light barrier of the drive motor is a pulse generator for the control unit. The light barrier <strong>and</strong> the incremental glass<br />

should be cleansed carefully with a cloth moistened with alcohol in intervals of 6 - 12 months. In "built in condition" of the<br />

glass you can also clean it with the help of a bristle brush, which has been soaked in alcohol. Clean the optical elements<br />

in between the glass <strong>and</strong> the light barrier carefully.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 139 of 213