[8.B1] OP39 Elevation: display as set point or true<br />

value<br />

(This option is only available for<br />

treadmills, not for a ladder-ergometer)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 113 of 213<br />

optional settings<br />

0 = Display of elevation as set point<br />

1 = Display of elevation as true value rsp. flashing set point report, if<br />

the set point was changed with the keys <strong>and</strong> has not yet been<br />

reached. (Report: current elevation angle; recommended defaultsetting)<br />

OP40 Speed: display as set point or true value 0 = Display of speed as set point (Report: target speed)<br />

1 = Display of speed as true value rsp. flashing set point report, if the<br />

set point was changed with the keys <strong>and</strong> has not yet been reached.<br />

(Report: current speed; recommended default-setting)<br />

OP41 Elevation control after STOP<br />

0 = Elevation maintains at the present level after pressing the STOP<br />

(This option is only available for button (recommended; default-setting)<br />

treadmills, not for a ladder-ergometer). 1 = Elevation gets down to 0 % after pressing the STOP button.<br />

The setting of option 41 to value 1 effects the<br />

safety of the subject <strong>and</strong> of third parties. A downmovement<br />

of the elevation after pressing STOP<br />

can result in a jumpy reaction <strong>and</strong> in dangerous<br />

injuries <strong>and</strong> contusions.<br />

OP46 Display <strong>and</strong> setting of elevation for<br />

h/p/cosmos strongly recommends to keep the<br />

st<strong>and</strong>ard setting of this option. Stop means Stop<br />

for all movements <strong>and</strong> must not result in an active<br />

action of the <strong>running</strong> <strong>machine</strong>. It is only allowed<br />

to change option 41 after a written documented<br />

instruction about the danger warnings <strong>and</strong> after a<br />

written confirmation of the customer/user.<br />

$ indicates: Elevation of ladder ergometer, st<strong>and</strong>ard: 75 degree,<br />

ladder ergometer<br />

(only available for ladder ergometers,<br />

not for treadmills)<br />

adjustable from 0 to 180 degree<br />

OP47 Set back of oil- <strong>and</strong> service interval After a service with lubricating the treadmill <strong>and</strong> performing all<br />

maintenance work the oil <strong>and</strong> service interval can be set back with<br />

option number OP47.